6σ质量管理理论在电化学发光法测定肿瘤标志物中的应用 被引量:10

The application of 6σ theory for quality management in the detction of tumor markers using electrochemiluminescence system
下载PDF
导出
摘要 目的探讨六西格玛(6σ)质量管理理论在电化学发光测定肿瘤标志物中的应用。方法根据Westgard等报道的方法计算σ值,分别选用以生物学变异确定的"合适"及"理想"允许总误差(TEa)为质量目标,不精密度(CV%)源于为该室室内质控累积CV%,不准确度(Bias%)源于卫生部临床检验中心肿瘤标志物室间质评结果。结果不同质量目标有不同的σ值及不同的质控方案。结论合理地采用质量目标,结合6σ质量管理理论指导质量控制方案的设计,适合于临床检验质量管理。 Objective To investigate the application of 6σ theory for quality management in the determination of tumor markers using electrochemiluminescence system.Methods σvalue was calculated by Westgard method,and the suitable and ideal TEa(allowable total error)were selected as the quality target based on the biological variation,and the imprecision derived from the internal quality control CV accumulated in our laboratory,and the inaccuracy derived from the results of tumor marker of external quality assessment in clinical laboratory center of ministry of health.Results The different quality target would possess the different σvalue and quality control procedure.Conclusion It is compatible for clinical laboratory to use the quality target reasonably and design the quality control procedure based on 6σ theory.
出处 《重庆医学》 CAS CSCD 北大核心 2010年第24期3342-3343,共2页 Chongqing medicine
关键词 六西格玛 允许总误差 生物学变异 质量控制 Tea biological variation quality control
  • 相关文献

参考文献13

二级参考文献36

  • 1李萍,刘小娟,徐克和,黄志刚.利用Westgard标准决定图判定测定方法性能[J].临床检验杂志,2006,24(1):69-70. 被引量:27
  • 2陈文祥,申子瑜,杨振华.临床检验分析质量指标的设定[J].中华检验医学杂志,2006,29(4):298-300. 被引量:51
  • 3段惠玲,李忠俊,滕本秀.应用“双质控法”、加强“即刻法”室内质控[J].重庆医学,2006,35(11):971-972. 被引量:9
  • 4Nevalainen D, Berte L, Kraft C, et al. Evaluating laboratory performance on quality indicators with the six sigma scale [ J]. Arch Pathol Lab Med, 2000, 124 (4) : 516-519.
  • 5Westgard JO, Darcy T. The truth about quality: medical usefulness and analytical reliability of laboratory tests [J]. Clin Chim Acta,2004,346(1) :3.
  • 6Westgard JO. The need for a syatem of quality standards for modern quality management[J]. Scand J Clin Lab Invest, 1999,59(7) :483.
  • 7Westgard AO. Internal quality control: Planning and implementation strategies[J]. Annals of Clinical Biochemistry,2003,40(6) :593.
  • 8Nevalainen D,Berte L,Kraft C,et al.Evaluating la-boratory performance on quality indicators with the six sigma scale[J].Arch Pathol Lab Med,2000,124(4):516-519.
  • 9Westgard JO,Westgard SA.The quality of laboratory testing today:an assessment of sigma metrics for analytic quality using performance data from proficiency testing surveys and the CLIA criteria for acceptable performance[J].Am J Clin Pathol,2006,125(3):343-354.
  • 10National Cholesterol Education Program Laboratory Standardization Panel.Current status of blood cholesterol measurement in clinical laboratories in the United States.Clin Chem,1988,34:193-201.

共引文献138

同被引文献69

  • 1王国民,陈伟.泌尿系统肿瘤治疗的进展与展望[J].肿瘤防治研究,2014,41(2):97-101. 被引量:25
  • 2王荫椿.肉毒毒素及其临床应用(综述)[J].中国神经免疫学和神经病学杂志,1996,3(2):124-127. 被引量:56
  • 3翟建新,焦连亭.浅谈我国室间质控评价计划在运行中存在的问题[J].实用医技杂志,2007,14(16):2189-2190. 被引量:1
  • 4Westgard JO,Carey RN,Wold S.Criteria for judging precision and accuracy in method development and evaluation[J].Clin Chem,1974,20(7):825-833.
  • 5Bland JM,Altman DG.Statistical methods for assessing agreement between 2 methods of clinical measurement[J].Lancet,1986,1(8476):307-310.
  • 6Clarke WL,Cox D,Gonder-Frederick LA,et al.Evaluating clinical accuracy of systems for self-monitoring of blood glucose[J].Diabetes Care,1987,10(5):622-628.
  • 7Parkes JL,Slatin SL,Pardo S,et al.A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose[J].Diabetes Care,2000,23(8):1143-1148.
  • 8Hillege HL,Janssen WM,Bak AA,et al.Microalbuminuria is common,also in a nondiabetic,nonhypertensive population,and an independent indicator of cardiovascular risk factors and cardiovascular morbidity[J].J Intern Med,2001,249(6):519-526.
  • 9Kjeldsen SE,Westheim AS,Os I.Prevention of cardiovascular events and diabetes with angiotensin-receptor blockers in hypertension:Life,Scope,and Value[J].Curr Hypertens Rep,2005,7(3):155-157.
  • 10CLSI.Clinical and Laboratory Standards Institut.EP27-P How to construct and interpret an error grid for diagnostic assays:approved guideline[S].Wayne,PA:,2009.