摘要 采用氢化物发生原子荧光法直接测定不同粒径大气颗粒物中As(Ⅲ)和As(Ⅴ)的含量.研究了还原剂用量、酸介质及其酸度、载气及屏蔽气流量和观测高度等对荧光强度的影响,探讨了共存离子对砷测定的干扰.在选定的最佳条件下,得到检出限为0.34μg/L,方法检出限为0.21μg/g,加标回收率为98.18%~102.54%,相对标准偏差为0.8%左右.用该方法测定大气颗粒物中不同形态的砷,操作简便,快速,灵敏度高. Concentration of atmospheric particles is one of the atmospheric pollution indicators.Heavy metals in the atmospheric particles can risk for human health in both direct and indirect way.Arsenic is one of the higher metal content in them.The inorganic compounds are far more toxic than their organic metabolites.In this paper,the hydride generation atomic fluorescence spectrometric method was employed to the determination of As(Ⅲ) and As(Ⅴ) in the Atmospheric particles.The amount of reducing agent,acid medium and its acidity,carrier gas and shield gas flow rate and observation height of the fluorescence intensity were investigated,and the interference experiment was carried out for concomitant elements.In the best conditions,the detection limit was 0.34 μg/L,the recovery ranged from 98.18%~ 102.54%,and the relative standard deviation was about 0.8%.The method was featured by easy operation,fast speed and it has been applied to the analysis of arsenic in the particles with satisfactory results.
出处 《河北大学学报(自然科学版)》 CAS 北大核心 2011年第1期42-47,共6页 Journal of Hebei University(Natural Science Edition)
基金 河北省科学技术研究与发展指导计划项目(07276724)
关键词 氢化物发生-原子荧光光谱法 大气颗粒物 形态分析 arsenics hydride generation-atomic fluorescence spectrometry atmospheric particles speciation analysis
  • 相关文献

参考文献13

  • 1CttAPPELL W R, BECK B D, BROWN K G. Toxocology profile For Arsenic[J]. Environ Health Perspect: 1997,1OS: 1060-1110.
  • 2MICHAEL P, LARRY K, BHALCHANDRA A. Induction of proliferative lesions of the uterus, testes, and liver in swiss mice given repeated injections of sodium arsenate: possible estrogenic mode of action[J]. Toxicol and Appl Pharma, 2000,166: 24-35.
  • 3GUO T, BAASNER J ,TSALEV D L. Fast automated determination of toxicologically relevant arsenic in urine by flow injection-hydride generation atomic absorption spectrometry[J]. Anal Chim Acta, 1997, 349: 313-318.
  • 4JERZY M, SAMUEL B A, HARKIRAT S D. Slurry sampling for hydride generation atomic absorption spectrometric determination of arsenic in cigarette tobaccos[J]. Analyst, 1997,122 : 539-542.
  • 5易江 许月英 等.氢化物发生原子吸收分光光度法测定蔬菜中的微量砷[J].光谱学与光谱分析,1987,8(2):67-70.
  • 6GUO Tie zheng, LIU Ming zhong, SCHRADER W. Use of flow injection hydride generation technique in non-dispersive atomic fluorescence spectrometry[J]. J Anal At Spectrom, 1992, 7: 1-8.
  • 7乔凤霞,孙汉文,闫宏远,锁然,李丽欣.氢化物发生原子荧光法同时测定丹参中痕量砷和锑[J].河北大学学报(自然科学版),2004,24(2):163-167. 被引量:13
  • 8唐姣荣,陆建平,童张法.HG-AFS测定大气颗粒物中砷铅汞含量[J].广西大学学报(自然科学版),2009,34(2):187-190. 被引量:15
  • 9PACEY G E, FORD. Arsenic speciation by ion-exchange separation and graphite furnace atomic-absorption spectrophotometry[J]. Talanta, 1981, 28.. 935-938.
  • 10BUDESINSKY B W. Arsenic colorimetry with silver diethyldithiocabamate[J]. Microchem J, 1979, 24: 80-87.

二级参考文献33

  • 1叶鹏,张剑波,陈辉,左澎.分光光度法测定大气中的过氧化物[J].中国环境监测,2005,21(1):43-47. 被引量:3
  • 2王湘硚.砷钼蓝萃取光度法测定硫磺中微量砷[J].理化检验(化学分册),1995,31(1):52-53. 被引量:3
  • 3陈柏林,谢君.毒物砷的测定及价态分析方法[J].四川农业大学学报,1995,13(2):242-246. 被引量:8
  • 4蔡卓,梁信源,钟伟明,韦韩亮.单扫描极谱法测定瓜子中痕量铅的研究[J].广西大学学报(自然科学版),2006,31(3):233-236. 被引量:6
  • 5陆建平,王益林,蒋桂海.罗丹明6G荧光猝灭法测定Hg^(2+)离子[J].广西大学学报(自然科学版),2007,32(1):31-34. 被引量:5
  • 6FAOUZIA E H,ANGEL M R, MIGUEL D G. Determination of total arsenic in .soft drinks by hydride generation atomic fluorescence spectrometry [ J ]. Food Chemistry, 2007,105 : 1195-1200.
  • 7WIFLADT A M, WIBETON G, LUND W. Determination of antimony in wine by hydride generation graphite atomic absorption spectrometry[J]. F J Anal Chem, 1997,357(1): 92-96.
  • 8LABORDA F, BOLEA E, LUAN R. Tubular electrolytic hydride generation for continuous and flow injection sample in atomic absorption spectrometry[J]. J Anal At Spectrom, 2000,15(3): 103 - 107.
  • 9MIERZWA J, ADELOJU S B, DHINDSA H S. Slurry sampling for hydride generation atomic spectrometric determination of arsenic in cigarette tobaccos[J]. Analyst, 1997,122: 539- 542.
  • 10SAYAGO A, BELTRAN R GOMEZ- ARIZA J L. Hydride generation atomic fluorescence spectrometry(HG- AFS as a sensitive detector for Sb( Ⅲ) and Sb( Ⅴ ) speciation in water[J]. J Anal At Spectrom, 2000,15(4): 423 - 428.

共引文献37

同被引文献61

引证文献6

二级引证文献31

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部