期刊文献+

DADE DIMENSION—RXL型全自动生化分析仪性能评价 被引量:1

Estimeation the character of autometed - biochemical analyzer of DIMENSION - RXL type The Department of exemination
下载PDF
导出
摘要 目的:评价美国杜邦DIMENSION—RXL全自动生化分析仪的工作性能。方法:用DADE试剂及定值血清对该仪器的准确度、精密度及交叉污染等进行了观察。结果:对包括酶活性及电解质在内的十二项待测物进行了测试,其结果均位于可信范围内。精密度观察:批内不精密度的CV%为0.07-1.45%,批间不精密度CV%为0.58-1.95%。交叉污染率为0-0.551%。结论:通过准确度、精密度和交叉污染实验观察,作者认为该仪器可以满足临床化学实验室的要求。 Objective: estimation the character of automated - biochemical analyzer made in USE by DuBanof type DIMENSION - RXL; Methods: Investigation the analyzer's accuracy, precision, liner and crosscontamination ect. used ty the DADE reagent and quantative serum; Results: There were twelve items havebeen test including the enzyme activity and the electrolyte ect. The result are all within the crdible range. Thewithin - run imprecision of this method (CV% ) was 0. 07 - 1. 45 %, day to day waso. 58 - 1. 95 %. the ratioof cross contamination was 0 - 0. 551 %; Conclusion: The results of the experiments showed that the analyzermay be satisfied demand clinical chemical Lab
出处 《医疗装备》 1999年第12期7-10,共4页 Medical Equipment
关键词 全自动 生化分析仪 准确度 精密度 交叉污染 Automated - biochemical analyzer, accuracy precision, cross - contanination
  • 相关文献

参考文献12

二级参考文献26

  • 1帅江平.供求平衡状态下的城市自组织过程[J].地理学报,1996,51(4):374-383. 被引量:9
  • 2李秀彬.全球环境变化研究的核心领域──土地利用/土地覆被变化的国际研究动向[J].地理学报,1996,51(6):553-558. 被引量:1724
  • 3约翰·冯·杜能(1826) 吴衡康译.孤立国同农业和国民经济的关系[M].商务印书馆,1986..
  • 4米歇尔·沃尔德罗普(1995) 陈铃译.复杂[M].三联书店,1997..
  • 5杨吾扬 梁进社.高能经济地理学[M].北京:北京大学出版社,1997..
  • 6克劳斯.迈因策尔 曾国屏(译).复杂性的思维:物质、精神和人类的复杂动力学[M].中央编译出版社,1999..
  • 7福建省统计局.《福建统计摘要2002》[Z].,..
  • 8福建省人民政府办公厅 福建省人民政府发展研究中心.《海峡两岸繁荣带示意图》[A]..《福建跨世纪战略研究》(上册)[C].福建人民出版社,1996..
  • 9黄泽民.《关于晋江城市化的思考》[A].陆学艺主编.论文集《晋江模式与农村现代化》[C].知识出版社,1995..
  • 10克鲁格曼.发展、地理学与经济理论[M].北京:北京大学出版社,2000..

共引文献375

同被引文献5

引证文献1

二级引证文献4

相关作者