期刊文献+

汞型活性碳纸富集-XRF法测定地质试样中痕量碘 被引量:11

Preconcentration of Trace Iodine with A Disk of〔Hg^(2+).ACP〕 and Determination by XRFA
下载PDF
导出
摘要 本文用特制的汞型活性碳纸富集地质样品中痕量碘并用X-射线荧光光谱进行测定。方法操作简便,样品测定限为0.25ppm,测定含I^-量25μg薄试样片(n=10),其相对标准偏差为3%。经地质标准样品验证,结果与推荐值符合。 The iodine is enriched with a disk of Hg2+ loaded active charcoal paper which possesses a strong absorptivity as well as a large capacity. The background of the disk is low, and no matrix correction needed. The collection procedure of I- on the disk can be achieved simply by placing the flask containing the solution and the disk on a vibrator for 2-3 hours. The RSD is 3% (I-25μg, n=10). The detection limit is 0.25ppm for the sample. Data obtained from the analysis of CRMs GSS4-7, GSD-10 and GSD-12 by this method agreed well with certified values.
作者 刘德慧
出处 《岩矿测试》 CAS CSCD 北大核心 1990年第3期175-178,共4页 Rock and Mineral Analysis
关键词 地质 样品 活性碳纸 XRF法 iodine determination, mercury loaded active charcoal paper, trace iodine preconcentration
  • 相关文献

同被引文献630

引证文献11

二级引证文献119

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部