摘要 目的探讨在没有质控品的情况下用新鲜全血对ADVIA2120血液分析仪进行室内质量控制的效果。方法每天用新鲜全血标本,在经溯源至参考方法的配套校准物校准的迈瑞BC-3000 Plus三分群血液分析仪及经其定值的新鲜血校准的ADVIA2120五分类血液分析仪上测定,比对两台仪器所测主要参数白细胞(WBC)、红细胞(RBC)、血红蛋白(HGB)、血小板(PLT)、红细胞比容(HCT)、平均红细胞体积(MCV),以此作为ADVIA2120五分类血液分析仪的室内质量控制。结果两台仪器PLT平均偏差均小于10%,其他参数平均偏差小于5%。均未超过美国CLIA′88能力比对检验的分析质量要求推荐的允许误差范围,但RBC及PLT有几次超出允许误差范围,按失控处理。结论同一个实验室内不同血液分析仪用新鲜血每日比对是控制质量的可行方法。 Objective To observe the effect of quality control for hematology analyzers with fresh whole blood without the control material.Methods Fresh whole blood samples were tested daily on the BC-3000 Plus analyzer which was calibrated with calibrate material recommend by manufacturer and on the ADVIA2120 analyzer,which was calibrated by the fresh whole blood.Results of such primary parameters as WBC,RBC,HGB,HCT,MCV and PLT were compared to establish the quality control method for the ADVIA 2120.Results The average differences of all parameters between the two analyzers were smaller than 5%,except for PLT,which was smaller than 10%.But the RBC and PLT were out of control,which were beyond the range of allowable error.Conclusion It was a recommendable method that laboratories carry out quality control with fresh whole blood daily comparison between different hematology analyzers.
出处 《重庆医学》 CAS CSCD 北大核心 2011年第28期2854-2855,2858,共3页 Chongqing medicine
关键词 血液 质量控制 血液分析仪 blood quality control hematology analyzer
  • 相关文献

参考文献8

二级参考文献15

  • 1国家质量技术监督局计量司.测量不确定度评定与表示指南[M].北京:中国计量出版社,2000.62-124.
  • 2Allison F S. A historical review of quality control in hematology [J ].American Technology, 1983,49: 625-632.
  • 3Watanabe K, Kawai Y, Takeuchi K. Defferenee of blood cell counts with reference blood cell counters in different makers Part Ⅱ [ J ]. Rinsho Byori, 1997,45: 185-189.
  • 4Peng Mingting,Syst J Int,1998年,8卷,79页
  • 5蔡辉,卫生统计学,1995年,1930页
  • 6ICSH.Reference method for the enumeration of erythrocytes and leukocytes.Clin Lab Haemat,1994,16:13l-138.
  • 7Lewis SM, England JM, Rowan RM, et al.Standardisation for haemocytometry. Lab pract, 1988,38:68-69.
  • 8ICSH HGB Expert Panel. Recommendations for reference method for haemoglobinometry in human blood and specifications for international haemiglobincyanide standard.J Clin Pathol,1996,49:271-274.
  • 9ICSH Cytometry Panel. Recommendation for reference method for packed cell volume. Lab Hematol, 2001, 7:148-170.
  • 10ICSH Cytometry Panel. Platelet counting by RBC/Platelet ratio method. Am J Clin Pathol, 2001,115: 460-464.