摘要
背景:目前市场上便携式医疗设备主要以锂电池供电为主,续航能力较弱,无法在野外长期工作。目的:为提高便携式医用设备供电电源的可靠性和续航能力,设计了高效、可靠、低成本的正弦波逆变器,详细阐述了设计方案,并制作了实验样机。方法:主电路由推挽升压变换器和单相逆变桥组成,采用高频变压器实现电压比调整和电气隔离,降低了噪声,提高了效率、减小了输出电压纹波。逆变器功率开关管采用了RCVD缓冲电路,确保逆变桥安全工作。控制部分采用集成脉宽调制芯片SG3524和正弦函数发生芯片ICL8038实现正弦波脉宽调制。结果与结论:实验样机体积减小到传统逆变器的1/4,效率达到86%,输出电压波形失真度小于5%,在复杂的工况下实现了220V/50Hz的市电输出。逆变器可以使用蓄电池作为电源,简单可靠、易于调试。
BACKGROUND:Currently,the medical device is toward smaller,portable development.Portable medical equipments are mainly powered by lithium batteries.Battery life is weak,unable to work in the wild.OBJECTIVE:In order to improve power reliability and endurance of portable medical devices,to design a high efficiency,credible and low cost sine-wave inverter.METHODS:High frequency inverter technology is selected.It contains high frequency transformer,which can fulfill voltage ratio regulation and electricity isolation,push-pull converter and single-phase bridge circuit as the main circuit,degrade noise,raise efficiency,reduce voltage ripple.Power switching tube adopts a RCVD cushion circuit to ensure safety working.Sine pulse width modulation as control form is selected by using integrated circuit SG3524 and ICL8038.RESULTS AND CONCLUSION:The experimental model machine reduces the volume to 1/4 of the traditional inverter,system efficiency reaches 86%.The experimental result indicates the distortion factor of the output voltage waveform is less than 5% and output is 220 V/50 Hz commercial power in complicated operating mode.The Inverter can be used as a power supply battery that is simple,credible,easy to debug.
出处
《中国组织工程研究与临床康复》
CAS
CSCD
北大核心
2011年第39期7347-7350,共4页
Journal of Clinical Rehabilitative Tissue Engineering Research