期刊文献+

一台便携式伽玛相机的数据采集系统设计

Development of data acquisition system for a portable compact gamma camera
原文传递
导出
摘要 介绍了一套用于便携式小型伽玛相机成像的高性能数据采集系统。为满足其便携式移动方便的要求,该数据采集系统的设计基于通用的USB接口,采用模块化的设计与系统集成。选用多通道数据同时采集的高速USB-ADC卡(DT9836)与LabVIEW软件开发平台,实现了高分辨小型伽玛相机多通道数据的同时采集、处理、图像显示与存储等功能。测试结果表明,该数据采集系统的采集速率可达200 K/s左右,能很好满足该便携式高分辨伽玛相机系统的成像要求。 In this paper, a high-performance data acquisition system is designed for a portable compact gamma camera. Based on a generic USB interface with modular design and system integration, it uses high-speed USB-ADC (DT9836) with NI LabVIEW sottware development platform to achieve simultaneous multi-channel data collection, processing, image display and record for the high-resolution gamma camera. Test results show that this data acquisition system can achieve a sampling rate of up to 200 K/s, well meeting the requirements of a high-resolution portable gamma camera.
出处 《核技术》 CAS CSCD 北大核心 2011年第11期851-855,共5页 Nuclear Techniques
基金 国家自然科学基金项目(10875162)(高分辨大视野的多针孔SPECT成像实验研究)
关键词 便携式 数据采集系统 伽玛相机 LABVIEW Portable, Data acquisition system, Gamma camera, LabVIEW
  • 相关文献

参考文献11

  • 1Blankenberg F G. Molecular imaging with single photon emission computed tomography. IEEE Eng Med Biol Mag, 2004, 23(4): 51-57.
  • 2Lewellen T. Recent developments in technology. Phys Med Biol, 2008, 53:R287 PET detector -R317.
  • 3Wojcik R, Majewski S, Kross B, et al. High spatial resolution gamma imaging detector based on 5" diameter R3292 Hamamatsu PSPMT. IEEE Trans Nucl Sci, 1998, 45(3): 487-491.
  • 4Wojcik R, Majewski S, Kross B, et al. Optimized readout of small gamma cameras for high resolution single gamma and positron emission imaging. Proc IEEE Nucl Sci Symp ConfRec, 2001, 3:1821-1825.
  • 5张猛蛟,漆玉金.高分辨小型伽玛相机位置读出电路研究[J].核技术,2007,30(7):629-632. 被引量:7
  • 6Qi Y J, Zhang M J, Zhao C L, et al. Performance Comparison of Subtractive Resistive Readout with Conventional Resistive Readout for a High-resolutionCompact Gamma Camera. Proc IEEE Nucl Sci Symp Conf Rec, 2007, M19-67:3725-3728.
  • 7金庆,漆玉金,赵翠兰,张猛蛟.高分辨小型伽玛相机的数据采集触发电路[J].核技术,2009,32(8):636-640. 被引量:1
  • 8刘蒙,漆玉金,赵翠兰,张雪竹.2×2阵列H8500光电倍增管简化读出电路的设计[J].核技术,2010,33(8):613-618. 被引量:3
  • 9Popov V, Majewski S, Weisenberger A G. Readout electronicsfor multianode photomultiplier tubes with pad matrix anode layout. IEEE Nucl Sci Symp Conf Rec, 2003, 3:2156-2159.
  • 10Popov V, Majewski S, Welch B. A novel readout concept for rnultianode photomultiplier tubes with pad. Nucl Instr Meth, 2006, A567(1): 319-322.

二级参考文献22

  • 1张猛蛟,漆玉金.高分辨小型伽玛相机位置读出电路研究[J].核技术,2007,30(7):629-632. 被引量:7
  • 2Francis G.Blankenberg,IEEE Engineering in Medicine and Biology Magazine,2004,23(4):51-57.
  • 3QI Y J,ZHANG M J,ZHAO C L,et al.Proc IEEE Nucl Sci Symp Conf Rec,2007,M19-67,3725-3728.
  • 4Anger H O.Rev Sci Instr,1958,29(1):27-33.
  • 5戴蓓蓓,钱志远.线性电子线路.合肥:中国科技大学出版社,2002.346-349.
  • 6Blankenberg F G. IEEE Eng Med Biol Mag, 2004, 23(4):51-57.
  • 7Lewellen T. Phys Med Biol, 2008, 53: R287-R317.
  • 8Datasheet: Hamamatsu Tube Assembly H8500. http:// www.hamamatsu.com.
  • 9Popov V, Majewski S, Weisenberger A G, et al. IEEE Nucl Sci Symp Conf Rec, 2001, 4:1937-1940.
  • 10Olcott P D, Talcott J A, Habte F, et al. IEEE Nucl Sci, 2005, 51(1): 21-27.

共引文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部