期刊文献+

Sound transducer calibration of ambulatory audiometric system utilizing delta learning rule

Sound transducer calibration of ambulatory audiometric system utilizing delta learning rule
下载PDF
导出
摘要 An efficient calibration algorithm for an ambulatory audiometric test system is proposed. This system utilizes a personal digital assistant (PDA) device to generate the correct sound pressure level (SPL) from an audiometric transducer such as an earphone. The calibrated sound intensities for an audio-logical examination can be obtained in terms of the sound pressure levels of pure-tonal sinusoidal signals in eight-banded frequency ranges (250, 500, 1 000, 2 000, 3 000, 4 000, 6 000 and 8 000 Hz), and with mapping of the input sound pressure levels by the weight coefficients that are tuned by the delta learning rule. With this scheme, the sound intensities, which evoke eight-banded sound pressure levels by 5 dB steps from a minimum of 25 dB to a maximum of 80 dB, can be generated without volume displacement. Consequently, these sound intensities can be utilized to accurately determine the hearing threshold of a subject in the ambulatory audiometric testing environment. An efficient calibration algorithm for an ambulatory audiometric test system is proposed. This system utilizes a personal digital assistant (PDA) device to generate the correct sound pressure level (SPL) from an audiometric transducer such as an earphone. The calibrated sound intensities for an audio-logical examination can be obtained in terms of the sound pressure levels of pure-tonal sinusoidal signals in eight-banded frequency ranges (250, 500, 1 000, 2 000, 3 000, 4 000, 6 000 and 8 000 Hz), and with mapping of the input sound pressure levels by the weight coefficients that are tuned by the delta learning rule. With this scheme, the sound intensities, which evoke eight-banded sound pressure levels by 5 dB steps from a minimum of 25 dB to a maximum of 80 dB, can be generated without volume displacement. Consequently, these sound intensities can be utilized to accurately determine the hearing threshold of a subject in the ambulatory audiometric testing environment.
出处 《Journal of Central South University》 SCIE EI CAS 2011年第6期2009-2014,共6页 中南大学学报(英文版)
基金 supported by the grant of the Korean Ministry of Education, Science and Technology (The Regional Core Research Program/Chungbuk BIT Research-Oriented University Consortium)
关键词 calibration audiometric system pure-tonal sound sound pressure level heating threshold delta learning rule 传感器校准 声音强度 规则系统 三角洲 门诊 听力 个人数字助理 声压级
  • 相关文献

参考文献24

  • 1NEWSBY H A, POPELKA G R. Audiology [M]. Englewood Cliffs: Prentice Hall, 1992:126-174.
  • 2DEBONIS D A, DONOHUE C R. Survey of audiology [M]. Bosten: Pearson Education, 2004: 77-106.
  • 3MARGOLIS R H, MORGAN D E. Automated pure-tine audiometry: An analysis of capacity, need, and benefit [J]. American Journal of Audiology, 2008, 17: 109-113.
  • 4KHANDPUR R S. Biomedical instrumentation, technology and applications [M]. New York: McGraw-Hill, 2005:463-485.
  • 5GIVENS G D, ELANGOVAN S. Internet application to tele-audiology-"nothin' but net" [J]. American Journal of Audiology, 2003, 12: 59-65.
  • 6YAO J, WAN Y, GIVENS G D. Using web services to realize remote hearing assessment [J]. Journal of Clinical Monitoring and Computing, 2010, 24:41 50.
  • 7SWANEPOEL D W, CLARK J, KOEKMOER D, HALL III J W, KRUMM M, FERRARI D V, MePHERSON B, OLUSANYA B O, MARS M, RUSSO 1, BARAJAS J J. Telehealth in audiology: The need and potential to reach underserved communities [J]. International Journal of Audiology, 2010, 49:195-202.
  • 8GIVENS G D, BLANAROVICH A, MURPHY T, SIMMONS S, BLACH D, ELANGOVAN S. Internet-based tele-audiometry system for the assessment of hearing: a pilot study [J]. Telemedicine Joumat and e-Health, 2003, 9(4): 375-378.
  • 9KRUMM M. Audiology telemedicine [J]. Joumat of Telemedicine and Telecare, 2007, 13: 224-229.
  • 10KRUMM M, RIBERA J, KLICH R. Providing basic hearing tests using remote computing technology [J]. Journal of Telemedicine and Telecare, 2007, 13: 406-410.