摘要 目的对1292快速生物检测法检测高压蒸汽灭菌效果进行评价,以解决传统生物检测法致骨科植入手术器械等待生物检测结果时间太长的问题。方法使用1292生物培养指示剂和第5类移动式化学指示卡作为灭菌过程验证装置,并比较两者的检测结果。结果 1292生物培养指示剂生物检测结果均为阴性,第5类移动式化学指示卡上的试剂爬行位置均进入规定合格窗口,两者检测合格率均为100%。结论 1292快速生物检测法效果可靠,所需时间短,操作简单,值得临床推广应用。 Objective Evaluations were conducted on the sterilization effect of high pressure steam in 1292 quick biological monitoring method so as to shorten the waiting time for the biological monitoring results of the traditional biological monitoring method of orthopaedic implants equipment. Methods Using 1292 organism cultivation indicator and category 5 mobile chemical indicator CARDS as the sterilization process ver bio on for ification device, the monitoring results logy cultivation biological monitoring in the card crawled into ACCEPT reagent both groups. Conclusion The 1292 were compared between the two dicators were negtive. Category 5 position(successful) window. The quick biological monitoring met convenient, thus deserves further clinical application. groups. Results The 1292 mobile chemical instructions monitoring pass were 100 hod is reliable, speedy and
出处 《解放军护理杂志》 2011年第23期74-75,共2页 Nursing Journal of Chinese People's Liberation Army
关键词 1292生物培养指示剂 第5类移动式化学指示卡 骨科植入手术器械 生物检测 1292 organism cultivation indicator category 5 mobile chemical indicator CARDS orthopaedicimplants surgical instruments biological monitoring
  • 相关文献

参考文献4

二级参考文献6

共引文献34

同被引文献9

  • 1梁树森,王华生,徐然.快读生物指示剂应用效果的观察[J].中华医院感染学杂志,2006,16(6):655-656. 被引量:3
  • 2中华人民共和国卫生部.消毒技术规范(2002年版)[S].北京:中华人民共和国卫生部,2002:15-43.
  • 3Albert H, Davies DJ, Woodson LP, et al. Biological indicators for steam sterilization: characterization of a rapid biological indi- cator utilizing Bacillus stearothermophilus spore-associated al- phmglucosidase enzyme [J]. J App Microbiol, 1998, 85 (5), 865- 874.
  • 4Vesley D, Langholz AC, Rohling SR, et al. Fluorimetric de tection of a Bacillus stearotherrnophilus sporebound enzyme, glucosidase, for rapid indication of flash sterilization failure [J]. Appl Environ Microbiol, 1992, 58 (2): 717-719.
  • 5Vesley D, Nellis MA, Allwood PB. Evaluation of a rapid readout biological indicator for 121℃ and 132 ℃ vacuumas sisted steam sterilization cycles [J]- Infect Control Hosp Ep- idemiol, 1995, 16 (5): 281 -286.
  • 6Alfa MJ, Olson N, DeGagne P, etal. Evaluation of rapid readout biological indicators for 132 degrees C gravity and 132 degrees C vacuum assisted steam sterilization cycles u- sing a new automated fluorescent reader [J]. Infect Control Hosp Epidemiol, 2002, 23 (7): 388-392.
  • 7褚金萍.快速生物监测阅读器在压力蒸汽灭菌效果监测中的应用[J].天津护理,2010,18(3):150-150. 被引量:6
  • 8任珊娜,谢秀庭.快速生物监测阅读器在供应室灭菌效果中应用[J].中华医院感染学杂志,2011,21(24):5170-5170. 被引量:6
  • 9肖艾萍,肖雪林,张桂兰.快速生物监测法在植入型手术器械灭菌监测中的应用[J].中国消毒学杂志,2012,29(2):158-159. 被引量:6

引证文献1

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部