摘要
目的:建立超高效液相色谱-串联质谱测定血浆中新型酪氨酸酶抑制剂UP302的方法,并用于研究UP302在大鼠、狗、猴和人血浆中的稳定性。方法:血浆中加入内标大豆苷元经2倍体积甲醇沉淀后进行分析。采用Hypersil Gold C18(50 mm×2.1 mm,1.9μm)色谱柱,流动相为甲醇(A)-5 mmol.L-1甲酸铵水溶液(B),梯度洗脱,梯度流速恒定为0.2 mL.min-1,柱温为30℃,整个分析时间为6 min。采用负离子电喷雾离子化电离源和选择反应监测模式进行检测。结果:在5~2000 ng.mL-1的浓度范围内,标准曲线线性关系良好(r=0.9998);血浆中UP302最低定量下限为5 ng.mL-1;本方法日内日间准确度在99.2%~107.3%,日内日间精密度均小于9.3%。结论:本方法灵敏度高,重现性好,操作简便,可用于血浆中UP302浓度的测定。
Objective:To establish an ultra -performance liquid chromatography -tandem mass spectrometry (UP- LC- MS/MS) method for the determination of tyrosinase inhibitor UP302 in plasma, and study the stability of UP302 in rat, dog, monkey and human plasma. Methods:The plasma samples were precipitated by methanol using daidzein as an internal standard. On Hypersil Gold C18column (50 mm × 2. 1 mm, 1.9 μm) , separation was achieved by a gradient mobile phase of methanol and 5 mmol · L^- 1 ammonium formate solution at a flow rate of 0. 2 mL · min^ - l at 30℃. The total analysis time was 6 min. Selection - reaction monitoring mode in negative - mode electrospray ionization was used to detect UP302 in plasma. Results:Calibration curve was linear in a range from 5 - 2000 ng · mL^- 1 with a correlation coefficient of 0. 9998. The lower limit of quantification was 5 ng · mL^- 1. The intra - and inter - day accuracy was between 99.2% and 107.3% , and precision was less than 9. 3%. Conclusion : The UPLC - MS/MS method is sensitive and reproducible. It can be used for the determination of UP302 in plasma.
出处
《药物分析杂志》
CAS
CSCD
北大核心
2012年第1期40-43,共4页
Chinese Journal of Pharmaceutical Analysis