期刊文献+

磁共振射频子系统主要参数对整体性能的影响 被引量:6

The Impact of Prime Parameters of MR RF-Subsystem on Functionality
下载PDF
导出
摘要 详细介绍了射频子系统的结构,主要参数及其对整体性能和临床诊断的影响,以及现有的相关射频技术及其发展。同时,结合采购过程中各大设备厂商提供的硬件参数,引用一些关键参数作了功能比较。 In this article,an introduction is given on the structure of RF subsystem,its main parameters and impact on the performance of MRI and RF techniques and its development.Furthermore,pivotal parameters provided by different suppliers are quoted and compared in functionality aspect.
作者 金玮 李斌
出处 《中国医疗器械杂志》 CAS 2012年第1期12-14,共3页 Chinese Journal of Medical Instrumentation
关键词 射频回路 射频放大器最大功率 射频通道 并行成像技术 加速因子 射频线圈 RF chain maxim output of RF amplifier RF channel parallel imaging accelerating factor RF coil
  • 相关文献

参考文献8

二级参考文献53

  • 1牛中奇,侯建强,周永军,王海彬,卢智远.生物电磁剂量学及人体吸收电磁剂量的数值分析[J].中国生物医学工程学报,2006,25(5):580-584. 被引量:15
  • 2[1]Lin FH,Wald LL,Ahlfors SP,et al.Dynamic magnetic resonance inverse imaging of human brain function[J].Magn Reson Med,2006,56(4):787-802.
  • 3[2]Glockner JF,Hu HH,Stanley DW,et al.Parallel MR imaging:a user's guide[J].Radiographics,2005,25(5):1279-1297.
  • 4[3]Weiger M,Pruessmann KP,Boesiger P.2D SENSE for faster 3D MRI[J].MAGMA,2002,14:10-19.
  • 5[4]Nakamura M,Ikeda N,Tsuboko T,et al.Clinical usage of parallel imaging of head three-dimensional MR image for analysis of VSRAD:a comparison of imaging statistical analytic results in SENSE acquisition method shortening acquisition time versus conventional acquisition method[J].Nippon Hoshasen Gijutsu Gakkai Zasshi,2006,62(10):1456-1462.
  • 6[5]Noebauer-Huhmann IM,Glaser C,Dietrich O,et al.MR imaging of the cervical spine:assessment of image quality with parallel imaging compared to non-accelerated MR measurements[J].Eur Radiol,2007,17(5):1147-1155.
  • 7[6]Frydrychowicz A,Bley TA,Winterer JT,et al.Accelerated time-resolved 3D contrast-enhanced MR angiography at 3T:clinical experience in 31 patients[J].MAGMA,2006,19(4):187-195.
  • 8[7]Zenge MO,Vogt FM,Brauck K,et al.High-resolution continuously acquired peripheral MR angiography featuring partial parallel imaging GRAPPA[J].Magn Reson Med,2006,56(4):859-865.
  • 9[8]Nael K,Fenchel M,Salamon N,et al.Three-dimensional cerebral contrast-enhanced magnetic resonance venography at 3.0 Tesla:initial results using highly accelerated parallel acquisition[J].Invest Radiol,2006,41(10):763-768.
  • 10[9]Nikolaou K,Kramer H,Grosse C,et al.High-spatial-resolution multistation MR angiography with parallel imaging and blood pool contrast agent:initial experience[J].Radiology,2006,241(3):861-872.

共引文献22