摘要
目的 评价基于外周动脉血管张力参数原理(peripheral arterial tone,PAT)的腕式睡眠监测仪Watch-PAT 200对阻塞性睡眠呼吸暂停低通气综合征(OSAHS)的诊断价值.方法 分别用Watch-PAT 200和多道睡眠监测图(PSG)对43例打鼾患者行整夜7h睡眠同步监测.由2名专业的分析人员采用双盲法分别对PSG结果进行人工分析;Watch-PAT睡眠监测结果由仪器自动分析.比较两种方法诊断OSAHS的一致性.结果 43例受试者平均((x)±s,以下同)年龄(42.2±12.2)岁,体质量指数(28.0±3.9) kg/m2.受试者PSG测得平均呼吸暂停低通气指数(AHI)为(34.9±29.9)次/h,PAT测得AHI为(36.0±29.2)次/h.对所有受试者PSG和PAT测得的AHI、最低血氧饱和度进行相关分析,相关系数分别为0.931和0.926,P值均<0.01.采用Bland-Altman法对两种检测方法进行一致性评价,95.3%的数据在95%一致性界限之内.以AHI≥5次/h为诊断OSAHS阳性标准,两种方法一致性检验Kappa=0.688,P=0.000,有统计学意义;Watch-PAT检测的灵敏度为94.7%,特异度为80.0%,误诊率为20.0%,漏诊率为5.3%,诊断符合率为93.0%.采用ROC曲线方法,分别以PSG测得AHI大于等于5或30次/h为阈值,Watch-PAT检测的灵敏度和特异度分别为82.6%和100.0%,以及95.0%和95.7%.结论 Watch-PAT 200可作为一种便捷工具诊断OSAHS.
Objective To assess the accuracy of a wrist-worn device (Watch-PAT 200) in the diagnosis of obstructive sleep apnea syndrome (OSAHS). Methods Forty-three adult subjects with suspected OSAHS simultaneously had a standard in-laboratory polysomnogram (PSG) and wore the WatchPAT 200 during a full-night recording.PSG sleep and respiratory events were scored according to standard criteria. The PSG recordings were blindly manually analyzed, while Watch-PAT data were scored automatically based on the algorithm developed previously.Results The mean age of the subjects was (42.2 ± 12.2) years ((x) ±s),and mean body mass index was (28.0 ±3.9) kg/m2.Mean PSG apnea hypopnea index (AHI) was (34.9 ± 29.9) events per hour,and mean PAT-AHI was (36.0 ± 29.2) events per hour.There was a significant correlation between PAT AHI and AHI by PSG (r =0.931,P 〈0.01 ).A Bland-Altman plot of PAT AHI and PSG AHI was also used to assess the accuracy of Watch-PAT 200.At lower levels of AHI,PAT tended to overestimate disease severity,while at higher levels of AHI,Watch-PAT underestimated severity.To assess sensitivity and specificity of Watch-PAT,constructed receiver operator characteristic curves using a variety of AHI threshold values (5,15 and 30 events per hour).For AHI ≥5events per hour as threshold value,the Watch-PAT diagnosing rate was 93%,and sensitivity as well as specificity were 94.7% and 80.0%.The misdiagnosis rate and missed diagnosis rate were 20.0% and 5.3%.Optimal combinations of sensitivity and specificity for the AHI threshold values ( 15 and 30 events per hour) were 82.6% and 100.0%,95.0% and 95.7% respectively.Conclusion The Watch-PAT 200 may offer an accurate,robust,and reliable ambulatory method for the detection of OSAHS,with minimal patient discomfort.
出处
《中华耳鼻咽喉头颈外科杂志》
CAS
CSCD
北大核心
2012年第2期112-116,共5页
Chinese Journal of Otorhinolaryngology Head and Neck Surgery