摘要
以HP-5毛细管柱为色谱柱,醋酸戊酯为内标物,采用氢火焰离子化检测器(FID),定量测定正硅酸乙酯含量。峰面积与浓度呈良好的线性关系,r=0.9999,精密度良好,回收率99.1%~101.2%。
Objective: To establish a method for the determination of tetraethyl orthosilicate with cap- illary column gas chromatography. Methods: Selecting amyl acetate as an internal standard, using HP-5 cap- illary column gas chromatography with FID detector to determine tetraethyl orthosilieate. Results: A better linear relation was found between the peak area and the concentration of tetraethyl orthosilicate, r = O. 9999, the precision was well, the recovery was 99.1%- 101.2%. Conclusion: This method for the determination of tetraethyl orthosilicate was simple, accurate and rapid.
出处
《云南化工》
CAS
2012年第1期48-50,共3页
Yunnan Chemical Technology
关键词
毛细管气相色谱法
内标法
正硅酸乙酯
capillary column gas chromatography
internal standard method
tetraethyl orthosilicate