摘要 目的 评价国产的高密度脂蛋白胆固醇(HDL-C)直接测定法在临床上应用的可行性。方法 将国产的直接测定法与硫酸葡聚糖镁(DS-mg^+2)沉淀法进行比较分析,观察其精密度,线性范围,准确度及部分干扰因素。结果 直接测定法的CV略小于DS-mg^+2法,与DS-mg^+2的相关性良好,Y=0.83x+0.26,r=0.92,HDL-C含量在2.30mmol/L时线性良好;黄疸。
出处 《福建医药杂志》 CAS 2000年第1期87-88,共2页 Fujian Medical Journal
  • 相关文献

参考文献2

  • 1[2]Warnick GR,et al.Dextran Sulfate-mg2+ Precipitation Procedure for guantitation of high density eipoprotein cholesterol proposed selected method.Chin Chem.1982,28:1379.
  • 2[3]Warnjck GR,Wood PD.National cholesterol education program recommendations for measurement of high-dinsity lipopotein diolesterol executive summary.Clin Chem,1995,41:1427~1433.

同被引文献27

  • 1姜宇海.化学修饰酶法测定高密度脂蛋白胆固醇[J].华北煤炭医学院学报,2006,8(1):34-34. 被引量:1
  • 2王抒,李培瑛,陈文祥,董军,蒋雷,国汉邦,唐蔚青,李红霞,李健斋.血清胆固醇标准参考物质的研制[J].中华医学检验杂志,1996,19(4):216-219. 被引量:7
  • 3赵水平.高密度脂蛋白基础与临床研究新进展[J].中华高血压杂志,2006,14(9):686-688. 被引量:8
  • 4李影林.高密度脂蛋白胆固醇测定.中华医学检验全书[M].北京:人民卫生出版社,1996:730-731.
  • 5Warnick GR-, Wood PDI National Cholesterol Education Program recommendations for measurement of high-density lipoprotein cholesterol: executive summary [ J ]. Clin Chem, 1995, 41 ( 10 ) : 1427-14-33.
  • 6Sampson ML, Aubry A, Csako G, et al. Tripl lipid screening: A homogeneous sequential assay for HDL-cholesterol, total cholesterol, and triglycerides[ J ]. Clin Chem ,2001 ,47 ( 3 ) :532-539.
  • 7Benlian P, Cartsier C, Hennache G, et al. Comparison of a new method for the direct and simultaneous assessment of LDL-and HDL-eholesterol with uhracentrifugation and established methods [J]. Clin Chem,2000,46(4) :493-505.
  • 8Wamick GR, Nauck M, Rifai N. Evolution of methods for measurement of HDL-cholesterol:from uhracentrifugation to homogeneous assays [ J ]. Clin Chem ,2001,47 ( 9 ) : 1579-1596.
  • 9Kimberly MM, Leary ET, Cole TG, et al. Selection, validation, standardization,and performance of a designated comparison method for HDL-cholesterol for use in the cholesterol reference method laboratory network [ J]. Clin Chem, 1999,45 ( 10 ) : 1803-1812.
  • 10Usui S, Nakamura M,Jitsukata K,et al. Assessment of between-instrument variations in a HPLC method for serum lipoproteins and its traceability to reference methods for total cholesterol and HDL- cholesterol[ J]. Clin Chem,2000,46( 1 ) :63-72.