摘要 目的探讨永久性心脏起搏器置入术的并发症及防治措施。方法采取回顾性分析方法,对1994年1月至2010年12月期间置入的397台次永久性起博器手术并发症进行总结。结果在397例置入永久性起搏器的患者中,发生起搏器并发症的共25例,其中囊袋血肿6例、囊袋感染3例、囊袋皮肤破溃2例、电极导线脱位3例、电极导线断裂1例、电极漏电1例、膈肌抽动1例、误穿锁骨下动脉并置入鞘管1例,气胸2例,起搏器综合征5例。结论永久性心脏起搏器置入术是一项成熟的技术,手术并发症相对较少,在本组病例中占6.29%,且没有危及生命的严重并发症。并发症主要发生在开展手术的前5年中,随着技术和器械的发展,并发症逐步减少,且积累了一定经验,值得同行借鉴。 Objective To discusses the complications and preventive measures in permanent cardiac pacing placement. Methods The complications in 397 cases of permanent cardiac pacemaker placement from January 1994 to December 2010 were analyzed with retrospective analysis method. Results Of 397 cases of permanent cardiac pacemaker placement, 6 cases had bursa bag hematoma, 3 cases had bursa bag infection, 2 cases had bursa bag skin burst, 3 cases had electrode wire dislocation, 1 case had electrode wire fracture, 1 case had electrode leakage, 1 case had diaphragmatic muscle twitch, 1 case had mistakenly wear subclavian artery and inserting sheath pipe, 2 cases had pneumothorax, 5 cases had pacemaker syndrome. Conclusion The placement of permanent cardiac pacemaker was a mature technology. In our study the incidence of complications is 6.29%, without life-threatening complications. Complications in surgery occurs mainly in the first five years, with the development of technology and equipment, complications is gradually reduced.
机构地区 解放军
出处 《中国心血管病研究》 CAS 2012年第3期211-212,共2页 Chinese Journal of Cardiovascular Research
关键词 心脏 起搏器 并发症 心律失常 Heart Pacemaker Complication Arrhythmia
  • 相关文献

参考文献3

二级参考文献6

共引文献10

同被引文献85

  • 1刘中梅,李淑敏,韩明华,郭涛,赵玲.永久心脏起搏器植入术中致命性心律失常分析[J].中华心律失常学杂志,2004,8(5):317-317. 被引量:2
  • 2任自文.起搏器心动过速[J].中国循环杂志,1994,9(7):430-434. 被引量:5
  • 3杨甫文,甄泽年,赵敏,陈贤明,陈辉,张雯.扁桃体切除术正常剂量利多卡因局部麻醉致毒性惊厥6例分析[J].临床耳鼻咽喉科杂志,2006,20(12):568-569. 被引量:2
  • 4易忠,郭继鸿.起搏电极导线脱位[J].中国心脏起搏与心电生理杂志,2006,20(4):367-368. 被引量:23
  • 5余培桢,王方正,吴海英,王锦志,张奎俊,田瑞国,鲁志民,朱克平,尹彦琳,路黎明.安置DDD起搏器143例的并发症分析[J].中国心脏起搏与心电生理杂志,1996,10(4):204-205. 被引量:6
  • 6Nielsen JC, Kristensen L,Andersen HR,et al. A randomized comparison of atrial and dual-chamber pacing in 177 consecutive patients with sick sinus syndrome:echocardiogTaphic and clinical outcome. J Am Coll Cardiol,2003,42: 614.
  • 7Milasinovic G, Sperzel J, Smith TW, et al. Reduction of RV pac- ing by continuous optimization Of the AV interval.Pacing Clin Electrophysio, 2006,29: 406.
  • 8Bank AJ, Gage RM, Burns KV. Right ventricular pacing, me- chanical dyssynchrony, and heart failure. Cardiovasc Transl Res, 2012,5: 219-223.
  • 9Sweeney MO,Hellkamp AS,Ellen Tlbogen KA, el al. Adverse effect Of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circula- tion, 2003,107:2932.
  • 10Sweeney MO, Bank A J, Nsah E, el al. Search AV extension and managed ventricular pacing for promoting atrioventricular conduc- tion (SAVE PACe)triM.Minimlzing ventricular pacing to reduce atrial fibrillation in sinus node disease. N Enql J Med,2007,357: 1000.

引证文献10

二级引证文献37

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部