摘要 目的 评价双腔起搏器睡眠功能的可靠性及临床应用效果。方法  18例因病窦综合征及完全性房室传导阻滞置入带有睡眠功能的双腔起搏器。睡眠时间设在 6~ 8小时。睡眠频率设为 4 0~ 5 0次 /min ,术后行Holter检查。结果 所有患者起搏器睡眠功能均正常工作 ,无 1例患者睡眠时有不适感。结论 双腔起搏器睡眠功能工作可靠 ,更趋生理化 ,并能节省电能 ,是起搏治疗时一项有实用价值的功能设置。 Objective To evaluate the efficacy of sleep function in patients treated with dual chamber pacer. Methods Eighteen patients affected by sick sinus syndrome and complete atrial ventricular block were implanted dual chamber pacemakers with sleep function. Sleep time was designed at 6 8 hours. The sleep rate was programmed at 40 50 bpm. All patients had received Holter examination after implantation of the pacer. Results The sleep function is working normally in all patients. Patients had no uncomfortable sensation during sleeping. Conclusion The sleep function is a reliable and useful new algorithm of dual chamber pacemaker.
出处 《中华心血管病杂志》 CAS CSCD 北大核心 2000年第1期48-49,共2页 Chinese Journal of Cardiology
关键词 起搏器 睡眠功能 心率 设置 Pacemaker, artificial Sleep Heart rate
  • 相关文献

参考文献10

二级参考文献140

共引文献84

同被引文献36

  • 1孙育民,王骏,严铭钰,王鸣和.右室心尖部起搏前后心功能水平的对比研究[J].中国心脏起搏与心电生理杂志,2004,18(6):418-418. 被引量:6
  • 2张建军,杨新春,胡大一,商丽华,汪爱虎,陈明.462例永久起搏器置入术并发症的相关因素分析及防治对策[J].中国心脏起搏与心电生理杂志,2004,18(6):456-457. 被引量:64
  • 3华伟,王方正,陈新.双腔起搏的一个新的功能——自动工作方式转换[J].中华心律失常学杂志,1998,2(1):75-76. 被引量:3
  • 4耿仁义,朱中林,华伟.永久性起搏器的植入手术.北京:人民军医出版社,2004,177—184.
  • 5Kusumoto FM, Goldsehlager N. Cardiac Pacing. N Engl J Med, 1996,334 : 89 - 97.
  • 6Fananapazir L,Epstein ND,Curiel RV, et al. Long- term results of dual - chamber (DDD) pacing in obstructive hypertrophic cardiomyopathy: evidence for progressive symptomatic and hemodynamic improvement and reduction of left ventricular hypertrophy. Circulation. 1994,90:2731 -2742.
  • 7Nishimura RA, Trusty JM, Hayes DL, et al. Dual - chamber pacing for hypertrophic cardiomyopathy: a randomized, double-blind, crossover trial. J Am Coll Cardiol, 1997,29:435 -441.
  • 8Hochleimer M, Hortnagl H, Fridrich L, et al. Long - term effiency of physiologic dual- chamber pacing in the treatment of end- stage idiopathic dilated cardiomyopathy. Am J Cardiol, 1992,70:1320 - 1325.
  • 9Linde C, Gadler F, Edner M, et al. Results of atrioventricular synchronous pacing with optimized delay in patients with severe congestive heart failure. Am J Cardiol, 1995,75:919 -923.
  • 10Gazeau S, Leclercq C, Lavergne T, et al. Effects of multisite biventricular pacing in patients with heart failure and intraventrieular conduction delay. N Engl J Med,2001,344: 873 -880.