摘要 飞行时间(TOF)采集和高清(HD)重建是提高正电子发射断层显像(PET)性能的最新技术。本研究通过阳性球状灶NEMA体型试验研究和临床病例分析,比较TOF+HD、HD、TOF+迭代重建和迭代重建4种方法对PET定量分析和图像质量的影响,以明确这些技术的实际应用价值,指导临床更好地选用相关技术和相应设备。 Time of flight(TOF) and high definition(HD) reconstruction are the latest technologies to improve the performance of PET.In this article,the image quality and quantitative analysis were compared among TOF+HD,HD,TOF+iterative and iterative reconstruction through NEMA body phantom studies and clinical case analysis.TOF improved quantification and detection of smaller lesions through providing higher contrast,while HD significantly increased the SUV and T/B values in all size.Good understanding of these techniques may be useful for selection of the corresponding PET/CT systems.
出处 《现代仪器》 2012年第4期31-35,共5页 Modern Instruments
基金 国家自然科学基金(81171370) 卫生部核医学重点实验室课题(KF200987)
关键词 正电子发射断层 飞行时间技术 高清重建 Positron emission tomography(PET) Time of flight(TOF) High definition(HD) reconstruction
  • 相关文献

参考文献19

二级参考文献136

共引文献71

同被引文献32

  • 1朱朝晖.PET-CT技术的发展现状与展望[J].现代仪器,2004,10(4):6-11. 被引量:30
  • 2朱朝晖.PET/CT在肿瘤学中的应用[J].现代仪器,2006,12(4):15-17. 被引量:17
  • 3朱朝晖,傅喆.PET采集处理方法对病灶检出能力和图像质量的影响[J].现代仪器,2006,12(4):21-24. 被引量:6
  • 4王学明,沈克涵.医学成像系统[M].北京:清华大学出版社.2006.
  • 5Frouin V, Comtat C, Reilhac A, et al. Correction of partialvolume effect for PET striatal imaging: fast implementationand study of robustness [J].J Nucl Med,2002,43:1715-1726.
  • 6Chen CH, Muzic RF Jr, Nelson AD, et al. Simultaneousrecovery of size and radioactivity concentration of smallspheroids with PET data [J].J Nucl Med,1999,40:118-130.
  • 7Go Akamatsu,Katsuhiko Mitsumoto,Kaori Ishikawa,Takafumi Taniguchi,Nobuyoshi Ohya,Shingo Baba,Koichiro Abe,Masayuki Sasaki.Benefits of Point-Spread Function and Time of Flight for PET/CT Image Quality in Relation to the Body Mass Index and Injected Dose[J].Clinical Nuclear Medicine.2013(6)
  • 8Iain Murray,Antonis Kalemis,Joe Glennon,Syed Hasan,Shuaib Quraishi,Thomas Beyer,Norbert Avril.Time-of-flight PET/CT using low-activity protocols: potential implications for cancer therapy monitoring[J].European Journal of Nuclear Medicine and Molecular Imaging.2010(9)
  • 9吴文凯.PET/CT诊断学[M].北京:人民卫生出版社,2009:16-32.
  • 10Moses WW.Time of flight in PET revisited[J]. IEEE Transactions on Nuclear Science,2003, 50(5).1325 -1330.