期刊文献+

用非参数统计方法评估检测体系的最低检出限 被引量:1

Evaluate the limit of detection of quantitative system by non-parametric statistics
下载PDF
导出
摘要 目的建立一套基于非参数统计原理的评估方案,用于空白标本检测结果为非正态分布、标准曲线为非直线线性的检测体系最低检出限(limit of detection,LoD)的确定。方法根据数据分布模型,以非参数统计方法确定空白标本检测结果的第95百分位数作为空白检测限(limit of blank,LoB),通过LoD=LoB+tβ×s低值的数学模型来评估LoD。结果以非参数统计方法建立的评估程序可通过验证程序证明该LoD值有效。结论对于空白标本检测结果不符合正态分布,标准曲线为非直线线性的资料可用非参数统计的评估方法来确定LoD。 Objective To set up a new procedure based on non-parametric statistics principle for determining the limit of detection (LoD), and estimate the LoD of an detection system which blank measurements is non-normal distribution or calibration curve is nonlinear. Methods By using theoretical distribution models and non-parametric approach to determine the 95th percentile of the blank measurements defines the limit of blank (LoB), then LoD is estimated as: LoD=LOB+tβx SDs. Results LoD detection system that was partly based on non-parametric statistics can be proved effective by proving program. Conclusion The proposed approach which is based on the theory of non-parametric theory appears useful for application in the field that blank distributions are nongaussian distribution and calibration curve is nonlinear.
出处 《中国现代医生》 2012年第28期108-110,共3页 China Modern Doctor
基金 浙江省教育厅2011年度科研计划项目(Y201122246)
关键词 最低检出限 非参数统计 非正态分布 Limit of detection Non-parametric statistics Non-normal distribution
  • 相关文献

参考文献10

  • 1International Organization for Standardization. Capability of detection- part 1 :terms and definitions( 11843-1 )1997 ISO Geneva,Switzerland.
  • 2International Organization for Standardization. Capability of detection- part 2 : methodology in the linear calibration case ( 11843-2 ) 2000 ISO Geneva, Switzerland.
  • 3International Organization for Standardization. Capability of detection- part 3:methodology for determination of the critical value for the re- sponse variable when no calibration data are used (11843-3)2001 ISO Geneva, Switzerland.
  • 4International Organization for Standardization. Capability of detection- part 4:methodology for comparing the minimum detectable value with a given value (11843-4) 2001 ISO Geneva, Switzerland.
  • 5Currie LA. Nomenclature in evaluation of analytical methods including detection and quantif- ication capabilities (IUPAC recommendations 1995 )[J]. Pure Appl Chem, 1995,67 (10) : 1699-1723.
  • 6Davenport JM,Schlain B. Testing claimed minimal detectable concentra- tions of in vitro medical diagnostic devices[J]. Clin Chem ,2000,46(10) : 1669-1680.
  • 7Brown EN,McDermott TJ,Bloch KJ,et al. Defining the smallest analyte concentration an immunoassay can measure[J]. Clin Chem, 1996,42(6 Pt 1 ) : 893-903.
  • 8Linnet K,Kondratovich M. Partly nonparametric approach for determin- ing the limit of detection[J]. Clin Chem,2004,50(4) :732-740.
  • 9冯仁丰.分析灵敏度(检测限)[J].上海医学检验杂志,2002,17(3):133-136. 被引量:67
  • 10Hahn GJ, Meeker WQ. Statistical intervals:A guide for practitioners[M]. New York :John Wiley & Sons, 1991:319.

二级参考文献3

  • 1[1]Davenport JM,Schlain B.Testing claimed minimal detectable concentrations of in vitro medical diagnostic devices[J].Clin Chem,2000,46:1669-1680.
  • 2[2]Diamandis EP,Yu J,Melegos DN.Ultrasensitive Prostate-specific antigen assaya and their clinical application[J].Clin Chem,l996,42:853-857.
  • 3[3]Stamey TA.Lower lin\mits of detection,Biological detection limits,functional sensitivity,or residual cancer detection? Sensitivity report on Prostate-specific antigen assays mislead clinicians[J].Clin Chem,1996,42:849-852.

共引文献66

同被引文献9

  • 1Zhang D,Fan G , Zhang J ,et al. Nonparametric methods for measure-ments below detection limit[ J]. Stat Med,2009,28?4):700-715.
  • 2Linnet K,Kondratovich M. Partly nonparametric approach for determi-ning the limit of detection[ J]. Clin Chem,2004,50(4):732-740.
  • 3Linnet K. Estimation of the limit of detection with a bootstrap-derived standard error by a partly non-parametric approach.Application to HPLC drug assays[ J]. Clin Chem Lab Med,2005 ,43(4):394-399.
  • 4Romdhani H, Lakhal-Chaieb L. On the association between variableswith lower detection Iimits[ J]. Stat Med,2011,30(26):3137-3148.
  • 5Desimoni E,Brunetti B. About estimating the limit of detection ofheteroscedastic analytical systems[ J]. Anal Chim Acta,2009 ,655 ( 1-2):30-37.
  • 6Zhang D,Fan C, Zhang J ,et al. Nonparametric methods for measure-ments below detection limit[ J]. Stat Med ,2009,28(4):700-715.
  • 7Lavagnini I,Badocco D,Pastore P,et al. Theil-Sen nonparametricregression technique on univariate calibration,inverse regression anddetection limits[ J]. Talanta,2011,15 (87): 180-188.
  • 8van den Heuvel E. Estimation of the limit of detection for quantalresponse bioassays[ J]. Pharm Stat,2011,10(3):203-212.
  • 9李飞,王伟佳,温冬梅,索明环,欧阳能良,严海忠,阚丽娟,张秀明.化学发光免疫法检测B型利钠肽的检测限值和功能灵敏度研究[J].检验医学与临床,2011,8(18):2185-2188. 被引量:4

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...