ChinaHeart左心辅助装置的动物在体存活实验 被引量:12

In vivo Survival Evaluation of the ChinaHeart Left Ventricular Assist Device
下载PDF
导出
摘要 验证磁悬浮离心式左心辅助装置—ChinaHeart VAD的在体血液相容性和可靠性,评估其对实验动物主要终末器官的影响。通过6例健康绵羊的在体存活实验进行评估。在心脏不停跳的情况下,装置入口插入左心尖,出口与降主动脉吻合。术前和术后做血常规、生化及血凝检查,术后连续监测动物存活及辅助装置运行状况。择期动物安乐死后,进行终末主要脏器的宏观和病理学检查。4只羊术后3~28 h内因麻醉或手术意外死亡;2只羊成功复苏并分别辅助20 d和38 d后施行安乐死。对于成功存活的2例实验动物,术后护理期间血常规、生化和血凝检查结果均在正常范围内,辅助装置无机械故障发生。尸检时1只羊(存活20 d)血泵泵体内发现一肉眼可见小块血栓沉积,其余无血栓形成;组织病理检查发现2只羊左肺贴近辅助装置部位均可见有小叶性肺炎,其余终末脏器无明显病变发生。ChinaHeart VAD在体辅助期间血液相容性能良好,对各主要终末脏器无显著影响,具有较好的可靠性;动物存活实验中麻醉呼吸管理和手术操作也至关重要。 An in vivo experimental investigation was performed in ovine model to evaluate hemocompatibility, reliability and end-organ effects of the ChinaHeart VAD, a magnetic left ventricular assist device. The device was implanted in 6 healthy male sheep. Under anesthesia, via left lateral thoracotomy, the inflow cannula was inserted into the left ventricular apex on beating hearts, and the outflow graft was anastomosed to the descending aorta. Routine hematologic and biochemical tests were performed preoperatively and postoperatively to evaluate peripheral organ functions. Pump operating parameters were recorded continuously until termination of the experiment. At the end of the experimental period the sheep was humanely killed, and the end-organs were examined macroscopically and histopathologieally. Two of the 6 animals survived for 20 and 38 days, the other four died of anesthesia or blood leakage at the inflow cannula or outflow graft within the first 3 to 28 hours. For the two survival animals, hematological and biochemical data showed no evidence of organ dysfunction during the experimental period. At autopsy, one small thrombosis formatted at the bottom housing in one pump (20- day survival sheep) , but no other thrombosis was found in the ventricular cavity of both two sheep. In addition, except lobular pneumonia was observed the left lung lobe adjacent to the pump, there were no histological changes in the other major end-organs. The ChinaHeart VAD system demonstrated excellent hemocompatibility and reliability for more than 20 days, and had no significant side effect on the major end-organs. In addition, aneshesia management and surgical procedures are crucial in animal survived experiment.
出处 《中国生物医学工程学报》 CAS CSCD 北大核心 2012年第5期736-741,共6页 Chinese Journal of Biomedical Engineering
基金 国家高技术研究发展计划(863计划)(2012AA041605) 北京市自然科学基金(2010311)
关键词 左心室辅助装置 ChinaHeart 磁悬浮血泵 在体实验 left ventricular assist device ChinaHeart magnetic levitation blood pump in-vivo test
  • 相关文献

参考文献23

  • 1徐晓佳,李帆.轴流式血泵的研究进展[J].北京生物医学工程,2005,24(5):394-397. 被引量:5
  • 2Timms D. A review of clinical ventricular assist devices [ J ]. Med Eng Phys, 2011, 33(9) : 1041 -1047.
  • 3Rogers JG, Butler J, Lansman SL, et al. Chronic mechanical circulatory support for inotrope-dependent heart failure patients who are not trans-plant candidates: results of the INTREPID trial [J]. J Am Coil Cardiol, 2007, 50:741 -747.
  • 4Nguycn DQ, Thourani VH. Third-generation continuous flow left ventricular assist devices [ J]. Innovations (Phila), 2010, 5 (4) : 250 -258.
  • 5Patel SM, Throekmorton AL, Untaroiu A, et al. The status of failure and reliability testing of artificial blood pumps [ J ]. ASAIO J, 2005, 51(4) : 440 -451.
  • 6Farrar D J, Bourque K, Dague CP, et al. Design features, developmental status, and experimental results with the Heartmate III centrifugal left ventricular assist system with a magnetically levitated rotor [ J ]. ASAIO J, 2007, 53 ( 3 ) : 310 315.
  • 7Wieselthaler GM, O Driscoll G, Jansz P, et al. Initial experience with a novel left ventricular assist device magnetically levitated rotor in a multi-institutional trial Heart Lung Transplant, 2010, 29 ( 11 ) : 1218 - 1225. clinical with a.
  • 8Morshuis M, Schoenbrodt M, Nojiri C, et al. DuraHeart magnetically levitated centrifugal left ventricular assist system for advanced heart failure patients [ J]. Expert Rev Med Devices, 2010, 7(2) : 173 -183.
  • 9张岩,胡盛寿,周建业,等.Fw型轴流泵的体外溶血与动物实验研究[J].中国胸心血管外壳临床杂志,2009,16(2):114-117.
  • 10吴广辉.轴流式心脏辅助装置及其发展现状[J].北京生物医学工程,2004,23(4):306-308. 被引量:5

二级参考文献90

  • 1蔺嫦燕.植入式左心辅助装置[J].国外医学(生物医学工程分册),1995,18(1):20-23. 被引量:2
  • 2蔺嫦燕,侯晓彤,吴广辉,李冰一,王景,潘红九.XZ-Ⅱ型轴流血泵的流场分析[J].北京生物医学工程,2005,24(6):405-409. 被引量:11
  • 3张岩,薛嵩,桂幸民,胡盛寿,孙寒松,朱晓东.运用三维数值模拟对人工心脏轴流血泵的设计和改进[J].中国生物医学工程学报,2007,26(1):35-41. 被引量:20
  • 4罗新锦,孙寒松,李楠,胡盛寿,许建屏,张岩.机械循环辅助对术后患者血浆游离血红蛋白的影响[J].中国胸心血管外科临床杂志,2007,14(3):173-176. 被引量:6
  • 5Chinn R, Dembitsky W, Eaton L, et al. Multicenter experience: prevention and management of left ventricular assist device infections. ASAIOJ, 2005, 51(4):461-470.
  • 6Haddad H, Elabbassi W, Moustafa S, et al. Left ventricular assist devices as bridge to heart transplantation in congestive heart failure with pulmonary hypertension. ASAIO J, 2005, 51 (4) :456-460.
  • 7Davies RR, Russo MJ, Hong KN, et al. The use of mechanical circulatory support as a bridge to transplantation in pediatric patients:an analysis of the United Network for Organ Sharing database. J ThoracCardiovasc Surg, 2008, 135(2) :421-427.
  • 8Anyanwu AC, Fischer GW, Plotkina I, et al. Off-pump implant of the Jarvik 2000 ventricular assist device through median sternotomy. Ann ThoracSurg, 2007, 84(4):1405-1407.
  • 9Zhang Y, Xue S, Gui XM, etal. A novel integrated rotor of axial blood flow pump designed with computational fluid dynamics. Artif Organs, 2007, 31(7) :580-585.
  • 10Zhang Y, Zhan Z, Gui XM, et al. Design optimization of an axial blood pump with computational fluid dynamics. ASAIO J, 2008, 54(2):150-155.

共引文献40

同被引文献121

引证文献12

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部