硅油眼眼轴长度的测量研究 被引量:2

Axial length measurement in silicone oil-filled eyes
下载PDF
导出
摘要 目的探讨硅油眼眼轴长度的测量方法,评价IOL Master测量硅油眼眼轴长度的准确性。方法选择2009年1月至2010年1月于我院行玻璃体切割联合硅油填充术患者26例(30眼),术前1d分别应用接触式A超和IOLMaster测量眼轴长度,用SRK/T公式计算IOL屈光度。于术后1个月,IOLMaster复测眼轴长度,比较三者眼轴长度的差异,并观察患者的屈光状态。结果应用接触式A超测得的眼轴长度为(24.01±2.62)mm,应用IOLMaster术前和术后所测得的眼轴长度分别为(24.23±2.91)mm、(24.38±3.18)mm,差异无统计学意义(χ2=2.684,P=0.261)。植入IOL且取出硅油的患者6例(7眼)根据术前接触式A超测算的预测屈光误差(mean predictive error,MPE)为(0.45±0.43)D,绝对屈光误差(mean absolute error,MAE)为(0.51±0.36)D,根据术前IOLMaster测算的MPE为(0.19±0.39)D,MAE为(0.33±0.26)D,根据硅油填充术后IOLMaster测算MPE和MAE分别为(0.11±0.35)D、(0.29±0.20)D,三者差异均无统计学意义(均为P>0.05)。结论 IOLMas-ter能准确测量硅油眼的眼轴长度,有无硅油不影响其测量结果。 Objective To discuss the measurements of axial length (AL) in sili- cone oil-filled eye ,and evaluate the accuracy of AL measured by IOL Master. Methods Thirty eyes of 26 patients with vitrectomy and silicone oil tamponade in Tianjin Medica/University Eye Centre were enrolled. IOL Master and contact A-scan was performed to measure the AL at 1 day before surgery, and intraocular lens power was calculated with the SRK/T formula. At 1 month after surgery, AL was measured by IOL Master again. The AL measurements by IOL Master ( before and after surgery) and contact Ascan were compared, and the refractive outcomes were followed up. Results AL measured by contact A-scan was(24.01 ±2.62) mm,and by IOL Master before and after surgery were(24.23 ± 2.91 )mm and( 24.38 ± 3.18 )ram, there was no statistical difference(x2 = 2.684 ,P = 0.261 ). The mean predictive error and mean absolute refractive error for 7 eyes of 6 patients underwent silicone oil removal and intraocular lens implantation were (0.45 ± 0.43) D and (0.51 ± 0.36) D by contact A-scan, which were (0. 19 ± 0.39) D and (0.33 ± 0.26) D by IOL Master before surgery, (0.11 ± 0.35 ) D and (0.29 ± 0.20) D by IOL Master after surgery, respectively, there was no statistical difference ( all P 〉 0.05 ). Couclusiou IOL Master can precisely measure the AL in silicone off-filled eye, and presence or absence of silicone oil tamponade does not influence its results.
出处 《眼科新进展》 CAS 北大核心 2012年第11期1079-1082,共4页 Recent Advances in Ophthalmology
关键词 眼轴 IOL MASTER 接触式A超 硅油眼 axial length IOL Master contact A-scan silicone oil-filled eye
  • 相关文献

参考文献11

  • 1李石磊,才娜.硅油填充状态下测算人工晶状体屈光度数准确性分析[J].眼科新进展,2008,28(1):52-54. 被引量:12
  • 2Bencic G,Vatavuk Z,Marotti M,Loncar VL, Petrie I,Andrijevic-Derk B,et al. Comparison of A-scan and MRI for the measure-ment of axial length in silicone oil-filled eyes[ J]. Br J Ophthal-mol,2009,93 {4) :502-505.
  • 3Parravano M,Oddone F,Sampalmieri M,Gazzaniga D. Reliabilityof the IOLMaster in axial length evaluation in silicone oil-filledeyes[J].^,2007,21(7) :909-911.
  • 4Sun HJ, Choi KS. Improving intraocular lens power prediction incombined phacoemulsification and vitrectomy in eyes with mac-ular oedema[ J]. Acta Ophthalmol,2011,89(6) :575-578.
  • 5Coleman J, Silverman R, Lizzi F, Rondeau MJ. Ultrasonographyof the eye and orbit[ M] ? Vol. 2. Philadephia:Lippincott Williams& Wilkins,2006:47-122.
  • 6Takei K,Sekine Y,Okamoto F,Hommura S. Measurement of ax-ial length of eyes with incomplete filling of silicone oil in the vit-reous cavity using X ray computed tomography[J].Br J Ophthalmol,2002.6(1) :47-50.
  • 7Dietlein TS,Roessler G, Liike C,Dinslage S, Roters S ’ Jacobi PC,et al. Signal quality of biometry in silicone oil-filled eyes usingpartial coherence laser interferometiy [ J]. J Cataract RefractSwty,2005 ,31(5) : 1006-1010.
  • 8Roessler GF,Huth JK,Dietlein TS,Dinslage S, Plange N,WalterV ,et al. Accuracy and reproducibility of axial length measure-ment in eyes with silicone oil endotamponade [ J ]. BrJ Ophthal-mo^2009(93(ll) :1492-1494.
  • 9Habibabadi HF, Hashemi H, Jalali KH, Amini A, Esfahani MR.Refractive outcome of silicone oil removal and intraocular lensimplantation using laser interferometry [ J ]. Retina,2005,25(2):162-166.
  • 10El-Baha SM, El-Samadonl A, Idris HF, Rashad KM. Intraopera-tive biometry for intraocular lens power calculation at siliconeoilremoval[ J]. Eur J Ophthalmol,2003 ,13(7) :622-626.

二级参考文献6

共引文献11

同被引文献34

  • 1赵明威,蒋宇振,黎晓新,姜燕荣.眼内硅油对屈光状态影响的理论推导与临床观察[J].眼科研究,2003,21(3):292-295. 被引量:12
  • 2陈艳,林季建.重硅油在玻璃体视网膜手术的应用[J].国际眼科纵览,2006,30(6):378-382. 被引量:4
  • 3Salouti R, Nowroozzadeh MH, Zamani M, Ghoreyshi M, Salouti R. Comparison of the ultrasonographic method with 2 partial coherence interferometry methods for intraocular lens power calculation[J]. Optometry,2011,82(3) :140-147.
  • 4Parravano M, Oddone F, Sampalmieti M, Gazzangiga D. Reliabil- ity of the IOLMaster in axial length evaluation in silicone oil-filled eye[J]. Eye,2007,21 (7) :909011.
  • 5Kunavisarut P, Poopattanalafl P, Intarated C, Pathanapitoon K. Accuracy and reliability of IOL master and A-scan immersion bi- ometry in silicone off-filled eyes [ J]. Eye, 2012,26 ( 10 ) : 1344- 1348.
  • 6Wang K,Yuan MK,Jiang YR,Bao YZ,Li XX. Axial length meas- urements before and after removal of silicone oil: a new method to correct the axial length of silicone-filled eyes for optical bi- ometry[J]. Ophthal Physiol Opt,2009,29(4) :449-457.
  • 7Bencic G, Vatavuk Z, Marotti M, Loncar VL, Petrie I ,Andrijevic- Derk B,et al. Comparison of A-scan and MRI for the measurement of axial length in silicone oil-filled eyes[ JJ. Br J Ophthal- tool, 2009,93 ( 4 ) : 502-505.
  • 8Coleman J,Silverman R,Lizzi F. Ultrasonography of the eye and orbit[M].Philadephia:Lippincott Williams &Wilkins,2006.47-122.
  • 9Hotta K. Refractive changes in silicone oil-filled pseudo-phakic eyes[J].Retina,2005,(02):167-170.
  • 10Elbendary A M. Predicted versus actual intraocular lens power in silicon-oil-filled eyes undergoing cataract extraction using automated intraoperative retinoscopy[J].Current Eye Research,2012,(08):694-697.

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部