方法决定图评价迈瑞BC-5300血细胞分析仪性能 被引量:2

Methods Decided Figure Evaluation Performance of Blood Cells Analyzer of Mindray BC-5300
下载PDF
导出
摘要 目的建立一种简单、准确的血细胞分析仪性能评价方法。方法依据CLIA’88规定的实验结果总允许误差(Tea%),用室间质量评价平均偏倚(Bias%)和室内质控变异系数(CV%),在Excel中用X,Y轴散点法作方法决定图(MD)、方法决定总图。结果5个评价项目X,Y数据点分别为:红细胞(O.95%,2.05%)、血细胞比容(1.44%,0.63%)、血红蛋白(o.96%,3.63%)、白细胞(1.95%,0.19%)、血小板(5.45%,5.93%);方法决定总图5个数据点分别为:(15.83%,34.15%),(24.00%,10.47%),(137.10%,51.83%),(13.00%,1.24%),(21.80%,23.74%)。结论方法决定图评价仪器性能简单、准确,值得临床推广。迈瑞BC-5300血细胞分析仪性能良好。 Objective To establish a simple and accurate evaluation method of blood cells analyzer's performance. So as to clinical inspection results provide reliable guarantee. Methods In excel,according to the provisions of the CLIA'88 to ex- perimental results total error allowed (Tea %), with the between laboratories quality evaluation average Bias (Bias%) and quality control in laboratory variation coefficient (CV%),we made XY axis scatter plot chart to get method decided (MD) figure and methods decided figure the assembly drawing. Results 5 evaluation projects data points respectively were: RBC (0.95%,2.05%),HCT (1.44%,0.63%),Hemoglobin (0.96%,3.63%/00) andWBC (1.95%,0.19%),platelet (5.45%, 5.93 %). Methods decided figure of the assembly drawing five grouse date points respectively were: (15.83 %, 34.15 % ), (24.00%,10.47%),(137.10%,51.83%),(13.00%,1.24%) and (21.80%,23.74%).Conclusion The method decided figure evaluating instrument performance is simple, accurate, and worth clinical application. Mindray BC-5300 blood cells an- alyzer has good performance.
出处 《现代检验医学杂志》 CAS 2012年第5期108-110,113,共4页 Journal of Modern Laboratory Medicine
基金 陕西省特色专业医学检验项目资助,项目编号2010-13-47.
关键词 方法决定图 血细胞分析仪 性能评价 Kevwords:methods decided figure blood cells analyzer performance evaluation
  • 相关文献

参考文献6

二级参考文献10

  • 1彭黎明,府伟灵,高行军.Sysmex SE-9000全自动血液分析仪性能的评价[J].临床检验杂志,2004,22(5):396-397. 被引量:10
  • 2[1]International Council for Standardization in Haematology.Guidelines for the evaluation of blood cell analyzer including those used for differential leukocyte and reticulocyte counting and cell marker application[J].Clin Lab Haematol,1994,16:157-174.
  • 3[2]Carter AB,Chavda N.Evaluation of the Sysmex SE9000 automated haematology analyzer[J].Sysmex J Int,1995,5:6-16.
  • 4England JM.Rowan RM,van assendemft OW,et al. Protocol for evaluation of automated blood cell counters[J]. Clin Lab Haemat. 1984,6:69 ~84.
  • 5O'Sullivan MB.Quality control of multichannel haematology analyzers:critique of current methods and the need for performance goals [J]. Clin Lab Haematol, 1990,12 suppl 1:3~12.
  • 6U.S. Department of Health and Social Services. Medicare, Medicaid, and CLIA Programs: Regulations implementing the Clinical Laboratory Improvement Amendments of 1988 (CLIA) . Final Rule[J]. Fed Regist,1992,57:7002-7186.
  • 7WESTGARD JO. A method evaluation decision chart (MEDx Chart) for judging method performance [J]. Clin Lab Science, 1995,8:277-283.
  • 8WESTGARD JO, CAREY RN, WOLD S. Criteria for judging precision and accuracy in method development and evaluation[J]. Clin Chem, 1974,20:825-833.
  • 9WESTGARD JO, BURNETr RW. Precision requirements for cost-effective operation of analytical processes [J]. Clin Chem, 1990,36:1629-1632.
  • 10谷小林,彭明婷.二级标准血液分析仪的质量控制与应用[J].临床检验杂志,2001,19(4):237-240. 被引量:34

共引文献19

同被引文献11

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部