摘要 通过对微盲孔支架管材和电抛光支架管材的拉伸对比实验及微盲孔支架和电抛光支架径向支撑力学性能的对比实验研究,测试结果显示微盲孔处理前后性能偏差很小,表明支架表面微盲孔结构基本不会影响支架的力学性能,证实了微盲孔支架的安全性,利于微盲孔支架的推广使用。 Through comparative tensile experiments of two kinds of stent tubes with micropores and electro polished and comparative radial strength experiments of two kinds of stents with micropores and electro polished,the results show that the deviation of mechanical properties of stent with micropores is very small in comparison with stent without micropores.This indicates that micropores structure on the stent surface almost does not have any effect on the mechanical performance of stent and proves that the stent with micropores is safe,which may help to promote the use of polymer-free stent direct drug-loading by micropores.
出处 《功能材料》 EI CAS CSCD 北大核心 2012年第24期3414-3416,共3页 Journal of Functional Materials
基金 国家"十一五"科技支撑资助项目(2006BAI03A13)
关键词 微盲孔支架 力学性能 电抛光支架 micropores stent mechanical property electro polished stent
  • 相关文献

参考文献8

  • 1Lanzer P, Sternberg K, Schmitz K P, et al. Drug-eluting coronary stent very late thrombosis revisited[J]. Herz, 2008, 33(5) :334-342.
  • 2Maisel W H. Unanswered questions-drug-eluting stents and the risk of late thrombosis[J]. N Engl J Med, 2007, 356(10) :981-984.
  • 3Nakazawa Gaku, Finn Aloke V, Kolodgie Frank D, et al. A review of current devices and a look at new technology: drug-eluting stents[J]. Expert Review of Medical De- vices, 2009 , 6(1):33-42.
  • 4Abizaid A, PhD M D, J Ribamar Costa Jr M D. New drug-eluting stents an overview on biodegradable and pol- ymer-free next-generation stent systems[J]. Circulation:Cardiovascular "Interventions, 2010,3 : 384-393.
  • 5Jan Steffelm, Franz R Eberli, Thomas F, et al. Drug-e luting stents--what should be improved? [J]. Annals of Medicine, 2008,40(4) : 242-252.
  • 6British Standard.. BS En14299. Non active surgical im- plants-paticular requirements for cardiac and vascular im- plants-specific requirements for arterial stents,2004, 12.
  • 7US Food and Drug Adminstration. Non-clinical tests and recommended labeling for intravascular stents and associ- ated delivery systems; guidance for industry and FDA Staff. US Department of Health and Human Services:Food and Drug Administration, Center for Devices and Radiological Health, January 13, 2005.
  • 8周承倜,董何彦.微型网状结构支架的力学性能研究[J].应用力学学报,2004,21(3):1-8. 被引量:9

二级参考文献8

  • 1[2]Stephan C. Schrader, & Rafael Beyar, Evaluation of Compressive Mechanical Properties of Endoluminal Metal Stents, Catheterization and Cardiovascular Diagnosis 44: 179-187, Wiley-Liss, Inc., 1998
  • 2[3]Regis Rieu, Paul Barragan, Catherine Masson, Jean Fuseri, Vincent Garitey, Marc Silvestri, Pierre Roquebert, and Joel Sainsous, Radial Force of Coronary Stents: A Comparative Analysis, Catheterization and Cardiovascular Interventions 46: 380-391, Wiley-Liss, Inc., 1999
  • 3[4]Paul Baragan, Regis Rieu, et al., Elastic Recoil of Coronary Stents: A Comparative Analysis, Catheterization and Cardiovascular Interventions 50: 112-119, 2000
  • 4[5]Duda Stephan H., Wiskirchen, Jakub, et a., Physical Properties of Endovascular Stents: An Experimental Comparison From the Laboratory, J. of Vascular and Interventional Radiology, JVIR, 11(5): 645-654, May 2000
  • 5[6]John A Ormiston, Simon R. Dixon, et al., Stent Longitudinal Flexibility: A Comparison of 13 Stent Designs Before and After Balloon Expansion, Catheterization and Cardiovascular Interventions 50: 120~124, 2000
  • 6[10]Shanley F.R., The Column Paradox, JAS, Vol. 13, 648, 1946
  • 7[11]Shanley F.R., Inelastic Column Theory, JAS, Vol. 14, No.5, 261~268, 1947
  • 8[13]Gellin S., The plastic buckling of long shells under pure bending, International J. of Solids and Structures, 16: 397~407, 1980