摘要 目的:通过理论、临床实践和实验室检测3种方法介绍非晶硒平板探测器伪影的产生,分析伪影产生的原因,探讨解决产生伪影的方法,延长非晶硒平板探测器的使用寿命。方法:通过实际工作和实验室检测伪影,针对同一幅图像分析其伪影产生的原因。结果:非晶硒平板探测器伪影产生的原因是点状脱膜所致,选择1000个胸部平片做临床试验,在原有的照相条件下增加照相剂量,可使散在小面积高密度伪影有所减少。结论:为了提高医学影像诊断的正确率,减少被检者的辐射剂量需更换新的平板探测器。 Objective: The ghost effect of amorphous selenium fiat panel detector (a-Se FPD) was studied by theory, clinical practice, and experiments. In order to increase the life of a-Se FPD, the causes of ghost effect as well as corresponding solutions were discussed. Methods: Ghost effects were detected through experiment and clinical practice, and the causes of ghost in a same image were analyzed. Results: The ghost effect in a-Se FPD is caused by point desquamation of selenium film. According to an analysis of a thousand chest images in clinic, an increasing of x-ray doses can reduce the number of dispersed small size ghosts. Conclusion: In order not to affect clinical diagnosis, replacing the a-Se FPD with a new one is necessary.
作者 李松强 张晔
出处 《中国医学装备》 2013年第2期21-23,共3页 China Medical Equipment
关键词 探测器 脱膜 环境条件 投照条件 Detector Point desquamation Ambient condition Exposure condition
  • 相关文献

参考文献16

二级参考文献31

共引文献115

同被引文献20

  • 1姜金凯.数字化成像多种应用方案促进现代化医院提升[J].世界医疗器械,2012,18(2):48-49.
  • 2国家药品监督管理局.YYT0741-2009数字化医用X射线摄影系统专用技术条件[S].国家药品监督管理局,2009-11-15.
  • 3国家药品监督管理局.YYT0796.1-2010数字X射线成像系统的曝光指数第1部分:普通X射线摄影的定义和要求[S].国家药品监督管理局,2010-12-27.
  • 4国家质量监督检验检疫总局.JJG1078-2012医用数字摄影(CR、DR)系统X射线辐射源[S].国家质量监督检验检疫总局,2012-06-18.
  • 5IPEM Report 91.Recommended standards for routine performance testing of diagnostic X-ray imaging systems[R]. IPEM, 2005.
  • 6国家药品监督管理局.YYT0590.1-2005医用电气设备数字X射线成像装置特性第1部分:量子探测效率的测定[S].国家药品监督管理局,2005-12-07.
  • 7国家药品监督管理局.Feed back comment for china CDC made the proposed DR QAguideline(draftversion6)[S].国家药品监督管理局,2012-9-14.
  • 8邹志孟,都美玲.Canon平板探测器数字化成像技术在急诊工作中的应用[J].世界医疗器械,2012,18(3):46-47.
  • 9Kasap S, Frey J B, Belev G, et al. Amorphous and polycrystal- line photoconductors for direct conversion flat panel X-ray image sensors[J].Sensors(Basel), 2011,11 (5): 5112-- 5157.
  • 10Okada Y, Sato K, Ito T, et al. A newly developed a-Se mam-mography flat panel detector with high sensitivity and low image artifact[C]. SPIE Medical Imaging. International Society for Op tics and Photonics, 2013 :86685V-86685V-9.

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部