国产心脏起搏器Qinming 2312S/M临床应用的研究 被引量:11

The study of the clinical application of domestic pacemaker Qinming 2312S/M
下载PDF
导出
摘要 目的观察国产Qinming 2312S/M型心脏起搏器的功能、应用特点及随访参数,并与进口起搏器相比较。方法对25例安装Qinming 2312S/M起搏器的患者进行术后即刻和术后3个月、6个月随访起搏参数变化进行记录,并与其它三组(各25例)分别安装进口单腔起搏器美敦力RES01、圣犹达2402或百多力Axios S的患者相应的起搏参数进行比较,随访患者病情以及起搏参数变化。结果 4组患者手术即刻和术后3个月、6个月的起搏参数(包括起搏阈值、阻抗和波幅)相互之间比较,无统计学差异(P>0.05)。结论国产Qinming 2312S/M起搏器的临床疗效不亚于同类型进口起搏器。 Objective To evaluate the clinical application of the single chamber pacemaker Qinming 2312S/M through comparing the pacing parameters among Qinming 2312S/M, Medtronic RES01, ST. Jude 2402, and Biotronic Axios S in follow-up. Methods Twenty-five cases of installing Qinming 2312S/M pacemaker patients undergoing operation immediately and after 3 months, 6 months follow-up pacing parameters were recorded, and the other three groups respectively installed single chamber pacemaker Medtronic RESO1, ST. Jude 2402 and Biotronic Axios S were also in follow-up. The pacing parameters in these four groups were compared. Results The difference in pacing parameters which included the pacing threshold, impedance and wave amplitude was not significant in the four groups during follow-up ( P〉0.05 ). Conclusion The clinical efficacy of pacemaker Qinming 2312S/M is not less than those imported pacemakers.
出处 《中国心脏起搏与心电生理杂志》 2013年第1期11-12,共2页 Chinese Journal of Cardiac Pacing and Electrophysiology
关键词 心血管病学 起搏器 起搏阈值 临床随访 Cardiology Pacemaker Pacing threshold Follow-up
  • 相关文献

参考文献3

二级参考文献11

  • 1[1]Griesbach L,Gestrich B.Restoration of circadian variation and physiologic rate behaviour through closed loop stimulation;RAPID study fiding[J].Prog Biomed Res,2001,6:83.
  • 2[2]Polley R,Malinowski K,Greco OT,et al.Validation of closed loop rate adaptive pacing based on right ventricular mechanics[J].Med Biol Eng Comp,1996,34(Suppl11,part 1):195.
  • 3Abraham WT,Young JB,Leon AR,et al.Effects of cardiac resynchronization on disease progression in patients with left ventricular systolic dysfunction,an indication for an implantable cardioverter-defibrillator,and mildly symptomatic chronic heart failure[J].Circulation,2004,110:2864.
  • 4Koplan BA,Kaplan AJ,Weiner S,et al.Heart failure decompensation and all-cause mortality in relation to percent biventricular pacing in patients with heart failure:Is a goal of 100% biventricular pacing necessary[J]? J Am Coll Cardiol,2009,53:355.
  • 5Lekkerkerker JC,van Nieuwkoop C,Trines SA,et al.Risk factors and time delay associated with cardiac device infections:Leiden Device Registry[J].Heart,2009,95:715.
  • 6Klug D,Balde M,Pavin D,et al.PEOPLE Study Group.Risk fac-tors related to infections of implanted pacemakers and cardioverterdefibrillators:results of a large prospective study[J].Circulation,2007,116:1349.
  • 7Bourge RC,Abraham WT,Adamson PB,et al.COMPASS-HF Study Group.Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure:the COMPASS-HF study[J].J Am Coll Cardiol,2008,51(11):1073.
  • 8Gasparini M,Anselme F,Clementy J,et al.ADVANCE CRT-D Investigators.Biventricular versus right ventricular antitachycardia pacing to terminate ventricular tachyarrhythmias in patients receiving cardiac resynchronization therapy:the ADVANCE CRT-D Trial[J].Am Heart J,2010,159(6):1116.
  • 9Crossley G,Boyle A,Vitense H,et al.Trial design of the clinical evaluation of remote notification to reduce time to clinical decision:the clinical evaluation of remote notification to reduce time to clinical decision(CONNECT)study[J].Am Heart J,2008,156(5):840.
  • 10Linde C,Abraham WT,Gold MR,et al.Randomized trial of cardiac resynchronizatian in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptoms[J].J Am Cull Cardiol,2008,52:1834.

共引文献9

同被引文献50

  • 1孟天宇,张勇,袁记行.国产心脏起搏器Qingming 2312S/M与进口心脏起搏器临床效果研究[J].中国医药科学,2014,4(9):183-185. 被引量:4
  • 2Yongbo XU,Yilong BAI,M.A.Meyers.Deformation, Phase Transformation and Recrystallization in the Shear Bands Induced by High-Strain Rate Loading in Titanium and Its Alloys[J].Journal of Materials Science & Technology,2006,22(6):737-746. 被引量:16
  • 3王方正,张澍,华伟,胡大一,陈新,蒋文平.全国心脏起搏器临床应用调查(2002至2005年)[J].中国心脏起搏与心电生理杂志,2007,21(3):272-275. 被引量:15
  • 4李洁,高树军,王冬梅.心理干预对起搏器携带者临床症状的影响[C]//第十三次全国心血管病学术会议论文集.哈尔滨:中华医学会,2011:209-210.
  • 5华为.临床适用心脏起搏技术[M].北京:人民卫生出版社,2012:2-5.
  • 6KennyT.心脏起搏器基础教程[M].郭继鸿,张玲珍,李学斌,主译.天津:天津翻译科技出版公司,2008:33-37.
  • 7Kennergren C, Bjurman C, Wiklund R, et al. A dingle-centre experience of over one thousand lead extractions[J]. Europace, 2009, 11(5): 612-617.
  • 8Biffi M, Bertini M, Saporito D, et al. Actual pacemaker longevity: the benefit of stimulation by automatic capture verification[J]. Pacing Clin Electrophsiol, 2010, 33(7): 873-881.
  • 9Lotze U, Fischer S, Hoers T, et al. Electrical performance and automatic capture characteristics of a 3.5 mm2 passive fixation lead during 1 year follow up[J]. Pacing Clin Electrophysiol, 2009, 32(8): 1050-1055.
  • 10王建安,项美香.国产心脏起搏器研究进展[EB/OL].http://www. 365heart. com/show/105227, shtml. 2015-1-22.