X线衍射法与焦磷酸消化法测定粉尘中游离二氧化硅含量的比较 被引量:5

A MatchDetermination of Free Silica by XRay Diffraction and Pyrophosphoric Acid Methods
下载PDF
导出
摘要 对粉尘中游离二氧化硅含量进行了X线衍射法和焦磷酸消化法的对比测定。结果表明 ,X线衍射法与焦磷酸消化法的结果之间有很好的相关性 ,其测定值在统计学上无显著性差异 (P >0 0 5 ) ,两方法测定值的相关回归方程是 ^Y =1 0 0 0 9X +2 2 6 2 6 ,r =0 9719。X线衍射法的相对偏差小于 10 % ,达到常规微量 (0 1~ 5mg)样本的分析要求 ; Free silica in geological exploration dust is measured by both Xray diffraction and pyrophosphoric acid methods.The results show good correlation between two methods with no significant difference.The correlation regression equation is =1 0009X+2 2626,r=0 9719.The relative deviation of Xray diffraction method is <10%.The precision of the proposed method meets the requirement of microanalysis.Xray diffraction method could eliminate the interference of other factors.
出处 《工业卫生与职业病》 CAS CSCD 2000年第5期268-270,共3页 Industrial Health and Occupational Diseases
基金 国家劳动部重点科学基金!(L -94-3 6)
关键词 粉尘 游离二氧化硅 X线衍射法 焦磷酸消化法 Respirable dust Free silica Xray diffraction Pyrophosphoric acid method
  • 相关文献

参考文献6

  • 1闫跃进,柯希妮.论地质勘探作业呼吸性粉尘监测与控制技术研究[J].劳动保护科学技术,1997,17(1):18-20. 被引量:7
  • 2闫跃进,呼吸性粉尘监测技术与防治方法,1998年,266页
  • 3程玉海,中华劳动卫生职业病杂志,1988年,6卷,1期,42页
  • 4陈国华,工业卫生与职业病,1986年,12卷,13期,139页
  • 5顾学箕,劳动卫生学,1984年,112页
  • 6中华人民共和国卫生部,车间空气中呼吸性矽尘卫生标准