溴化镧探测器温度特性研究 被引量:3

Study on Temperature Characteristics of LaBr3 Detector
下载PDF
导出
摘要 通过开展溴化镧探测器温度实验,研究了温度变化对溴化镧探测器性能的影响。研究表明,在-10℃50℃范围内,溴化镧探测器的能量分辨率变化不超过±5%,峰总比变化不超过±6%,不会影响探测器的使用。但是道址随温度变化明显,需要在环境温度变化时做好稳谱。 We report on the research of temperature influence on the performance of LaBr3 detector. The result of temperature experiments show that energy resolution of LaBr3 detector changed within ± 5 % and peak - to - total ratio of LaBr3 detector changed within ± 6% from - 10℃ to 50℃, and the temperature changes during the experiment had no effect on the use of LaBr3 detector. The channel of peak position changed obviously, thereby spectrum stabilization had to be done with the change of ambient temperature.
作者 张燕 罗明
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2013年第2期188-190,共3页 Nuclear Electronics & Detection Technology
关键词 溴化镧探测器 能量分辨率 峰总比 道址漂移 LaBr3 detector energy resolution peak - to - total ratio channel drift
  • 相关文献

参考文献5

  • 1高鑫,何元金.LaBr_3:Ce^(3+)闪烁晶体研究进展[J].核电子学与探测技术,2010,30(1):5-11. 被引量:29
  • 2ESPEC高低温交变湿热实验箱手册[z].爱斯佩克环境仪器有限公司.
  • 3苏秀彬,刘义保,刘洋,杨波,魏强林,张煜莉.LaBr_3:Ce(5%)闪烁探测器的MC研究[J].核电子学与探测技术,2010,30(9):1215-1219. 被引量:11
  • 4光电倍增管基础及应用[z].北京滨松光子技术股份有限公司,2010:67-70.
  • 5Saint - Gobain Crystals, BrilLance 380 data sheet, ht- tp://www, detectors, saint - gobain, com/uploaded- Files/SGdeteetors/Documents/Product_ Data_ Sheets/ BrilLanCe380 - data - sheet, pdf.

二级参考文献27

  • 1谢建军,杨培志,廖晶莹.卤化镧系LnX_3(Ce)闪烁晶体的研究进展[J].无机材料学报,2005,20(3):522-528. 被引量:18
  • 2Van Loef EVD, et aI. High-energy-resolution scintillator: Ce^3+ activated LaCl3 [J]. Appl Phys Lett, 2000, 77(10) :1467.
  • 3Van Loef EVD, et al. scintillation properties of Lacl3 :Ce crystals: fast, efficient, and high-energy resolution scintillators[J]. IEEE Trans Nucl Sci, 2001, NS248 (3): 341.
  • 4Van Loef EVD, et al.High-energy-resolution scintillator: Ce activated LaBr[J]. Appl Phys Lett, 2001, 79(10) : 1573.
  • 5Van Loef EVD, et al. Scintillation properties ol LaBr: Ce crystals: Fast, efficient and high-energy-resolution scintillators[J]. Nucl Instrurn Methods, 2002, A486:254.
  • 6Alan Owens, et al. The hard X-ray response of Ce-doped lanthanum halide seindllators[J]. Nucl. Instr. and Meth. , 2007, A574.158-162.
  • 7R. Pard, et al. LaBr3 :Ce scintillation gamma camera prototype for X and gamma ray imaging[J]. Nucl. Instr. and Meth. A(2007), doi:10. 1016/ j. nirna. 2007. 01. 111.
  • 8Surti S, et al. Investigation of lanthanum scintillators for 3-D PET[J]. IEEE Trans Nucl. Sci , 2003, 50(3) :348-354.
  • 9Valentine J D, et al. The light yield nonproportionality component of scintillator energy resolution[J]. IEEE Trans Nucl. Sci. , 1998, 45:512- 517.
  • 10F. Quarati, et al. X-ray and gamma-ray response of a 2" × 2" LaBr3 : Ce scintillation detector[J]. Nucl. Instr. and Meth. , 2007, A574:115-120.

共引文献35

同被引文献17

引证文献3

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部