期刊文献+

紫外分光光度法对二氧化氯含量的测定 被引量:3

MEASUREMENT OF CHLORINE DIOXIDE CONTENT BY ULTRAVIOLET SPECTROPHTOMETRY
下载PDF
导出
摘要 采用紫外分光光度法,于二氧化氯有特征吸收峰的 260nm波长下测定溶液中二氧化氯含量。对二氧化氯含量<300 mg/L的溶液,测定结果的相对标准偏差≤5.09%。样品溶液pH值在5~10与放置时间≤50 min时,对测定结果影响不大。 Ultraviolet spectrophotometry was used to measure chlorine dioxide content in solution at wavelength of 260 nm which is the characteristic absorption peak of chlorine dioxide. The relative standard deviation of measured results of the solution containing chlorine dioxide ≤ 300mg / L was ≤5.09%. When the pH value of the sample solution was 5-10 and the solution was laid aside for ≤5 min, there was no significant influence on measured results.
出处 《中国消毒学杂志》 CAS 2000年第3期178-180,共3页 Chinese Journal of Disinfection
关键词 二氧化氯 浓度 紫外分光光度法 含量 chlorine dioxide measurement of concentration ultraviolet spectrophotometry
  • 相关文献

参考文献3

  • 1Watanabe T.Spectrophotometric determination of chlorine dioxide with Cresol Red. Bunseki Kagaku,1994; 40(8): 373.
  • 2Watanabe T. Fluorophotometric determination of chlorine dioxide with chromotropicacid. Analytical Science, 1992; 8(2): 207.
  • 3WatanabeT.Colorimetricdeterminationofchlorinedioxidewithm-aminobenzylalcohol.AnalyticalScience,1994;10(1):35. (1999-09-10收稿1999-11-15修回)