摘要
铬(Ⅵ)对血红蛋白模拟酶催化荧光体系具有强烈的猝灭作用,据此建立了一种酶催化分光荧光法测定铬?的新方法。研究了溶液酸度、L-酪氨酸浓度、血红蛋白浓度、H2O2浓度及反应时间等因素对体系的影响。在pH 10.4的NH3.H2O-NH4Cl缓冲溶液中,当L-酪氨酸、血红蛋白和H2O2的浓度分别为1.4×10-4 mol/L、1.0×10-6 mol/L和3.5×10-5 mol/L时,测定铬(Ⅵ)的线性范围为2.0×10-6~1.0×10-4 mol/L,检出限为1.1×10-8 mol/L。1 000倍NO3-、SO42-、Na+、K+、Cl-、Br-,300倍PO43-、Al 3+、NH4+,50倍Mn2+、Mg2+、Fe2+、Cu2+,1倍Fe3+对铬(Ⅵ)的测定没有干扰。干扰较大的Fe3+,可加入过量的柠檬酸掩蔽。对浓度为4.8×10-5 mol/L的铬(Ⅵ)进行11次平行测定,其相对标准偏差为2.8%。该法已成功地应用于环境水样中铬(Ⅵ)含量的测定。
According to strong quenching effect of chromium on the catalytic fluorescence system of hemoglobin analog enzyme,a novel enzymatic catalytic spectrofluorimetry for determining chromium(Ⅵ) was established.The influence of acidity,L-tyrosine concentration,hemoglobin concentration,H2O2 concentration and reaction time on the system was studied.In NH3·H2O-NH4Cl buffer solution of pH 10.4 and the presence of 1.4×10-4 mol/L L-tyrosine,1.0×10-6 mol/L hemoglobin and 3.5×10^-5 H2O2,the linear range for the determination of chromium(Ⅵ) was 2.0×10-6-1.0×10-4 mol/L and the detection limit was 1.1×10^-8 mol/L.1000 times of NO3-,SO4^2-,Na+,K+,Cl^-and Br-,300 times of PO43-,Al3+,NH4+ and 50 times of Mn2+,Mg^2+,Fe^2+,Cu^2+ and 1 time of Fe3+ did not interfere in the determination of chromium(Ⅵ).Fe3+ which had more interference could be masked by adding an over dose of citric acid.The RSD was 2.8% for 11 parallel determination of 4.8×10^-5 mol/L chromium(Ⅵ),.The method was successfully applied for determining chromium(Ⅵ) in the environment water.
出处
《冶金分析》
CAS
CSCD
北大核心
2013年第5期55-58,共4页
Metallurgical Analysis
基金
河南省基础与前沿技术研究计划资助项目(112300410013)
关键词
铬(Ⅵ)
酶
催化
分光荧光法
血红蛋白
环境水样
chromium(Ⅵ)
enzyme
catalysis
spectrofluorimetry
hemoglobin
environment water