期刊文献+

PdI混合源剂量学参数的蒙特卡罗模拟

Monte Carlo Simulation of Dosimetric Parameters for Hybrid PdI Source in Brachytherapy
下载PDF
导出
摘要 依据美国医用物理学家协会(AAPM)在TG-43U1推荐的剂量学参数计算方法,针对671l型(3M)125I短程治疗源,用蒙特卡罗方法计算剂量率常数,并与相关数据进行了比较。在此基础上,计算研制的新型125I-103Pd混合源的剂量学参数,如剂量率常数、径向剂量函数和各向异性函数,计算结果对混合源的实验研究和应用具有一定指导意义。 According to dose calculation formula recommended by AAPM TG-43U1, dose rate constant of Model 6711 125I brachytherapy source was calculated by Monte Carlo meth- od. The calculation results were in good agreement with TG-43U1. Then, dose rate con- stant,radial dose function and anisotropy function of new hybrid PdI source were calculated by Monte Carlo method. Empiric equations were obtained for radial dose function.
出处 《同位素》 CAS 2013年第3期148-151,共4页 Journal of Isotopes
关键词 蒙特卡罗方法 剂量率常数 径向剂量函数 各向异性函数 Monte Carlo method dose rate constant radial dose function anisotropy func-tion
  • 相关文献

参考文献3

二级参考文献15

  • 1NATH R,ANDERSON L L,LUXTON G,et al.Dosimetry of interstitial brachytherapy sources:recommendations of AAPM radiation therapy committee task group No43[J].Medical Physics,1995,22:209-234.
  • 2RIVARD M J,COURSEY B M,DEWERD L A,et al.Update of AAPM task group No43 report:A revised AAPM protocol for brachytherapy dose calculations[J].Medical Physics,2004,31:633-674.
  • 3HUBBELL J H,SELTZER S M.Tables of Xray mass attenuation coefficients and mass energy-absorption coefficients from 1 kev to 20 MeV for elements Z= 1 to 92 and 48 additional sub stances of dosimetric interest,NISTIR 5632[R].Japan:NIST,2004.
  • 4BRIESMEISTER J F.MCNP-A genernal Monto-Carlo N-particle transport code (version 4B),LA-12625-M Manual 4B[R].USA:LANL,1997.
  • 5Rivard M J,Coursey B M,DeWerd L A, et al.2004 Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations[].Medical Physics.2004
  • 6Ali S,Meigooni S B,Awan N S, et al.Updated Solid WaterTM to water conversion factors for 125I and 103Pd brachytherapy sources[].Medical Physics.2006
  • 7Murphy M K,Piper R K,Greenwood L R, et al.Evaluation of the new cesium-131 seed for use in low-energy X-ray brachytherapy[].Medical Physics.2004
  • 8Meigooni A S,Dini S A,Awan S B, et al.Theoretical and experi- mental determination of dosimetric characteristics for ADVANTAGE Pd-103 brachytherapy source[].Applied Radiation and Isotopes.2006
  • 9Rivard M J.A discretized approach to determining TG-43 brach- therapy dosimetry parameters: Case study using Monte Carlo calcu- lations for the MED3633 103Pd source[].Applied Radiation and Isotopes.2001
  • 10Reniers B,Vynckier S,Scalliet P.Dosimetric study of a new palla- dium seed[].Applied Radiation and Isotopes.2002

共引文献7