容积调强旋转放疗的计划验证通过率对多叶准直器位置误差的灵敏度 被引量:19

The sensitivity of patient-specific VMAT QA to MLC positioning errors
原文传递
导出
摘要 目的 针对容积调强旋转放疗技术(VMAT),分析患者计划验证通过率对多叶准直器(MLC)位置误差的灵敏度.方法 选取6例双弧VMAT计划,引入MLC位置误差(±0.5 mm、±1 mm和±2 mm),模拟VMAT执行过程中MLC可能出现的系统误差,包括MLC射野宽度误差和MLC整体偏向一侧的误差.每个病例有10个计划,1个原计划和9个带有误差的新计划.利用螺旋形半导体探测阵列(ArcCheck)进行验证测量,得到每个病例原计划和新计划的剂量分布.采用绝对剂量结合等剂量距离差别的计算方法,以原计划计算剂量分布为参考,分别得到每个计划的通过率.结果 当评价指标为3%/3 mm时,6例原计划的平均通过率为96%,带有+1 mm、+2 mm、-2 mm射野宽度误差的计划和2 mm MLC整体偏向一侧误差的计划平均通过率下降8.8%、15.5%、6.1%和7.9%,这些MLC位置误差通过计划验证可以检测到,其他MLC位置误差对计划通过率影响小,无法检测到.2%/2 mm评价指标较3%/3 mm对MLC位置误差更敏感.结论 对于1 mm以内的MLC位置误差,VMAT计划的验证对其不敏感.为保证VMAT计划执行的准确性,需要针对MLC做专门的质量控制. Objective To evaluate the sensitivity of patient-specific volumetric-modulated arc therapy (VMAT) quality assurance (QA) to minor multileaf collimator (M LC) positioning errors.Methods Systematic multileaf collimator (MLC) positioning errors (+0.5 mm,+ 1 mm and +2 mm) were introduced into the clinical VMAT patient plans with 2 types of MLC positioning errors:systematic MLC gap width errors and systematic MLC shift errors for 6 cases,including 3 cases with prostatic cancer and 3 cases with nasopharyngeal cancer.The planar dose distributions of the original and modified plans were measured using ArcCheck array.The coincidence between the measured results and the calculated results was evaluated using both absolute distance-to-agreement (AD-DTA) analysis with 3%/3 mm and 2%/2 mm criteria.Results The average passing rate of the 6 original VMAT plans was 96.0% with the ADDTA criteria of 3%/3 mm which was commonly adopted in clinical practice.For the MLC gap width errors of + 1 mm,+2 mm,and-2 mm and the MLC shift errors of 2 mm,the drop levels in average passing rate with the AD-DTA criteria of 3%/3 mm were 8.8%,15.5%,6.1% and 7.9%,respectively.The + 2 mm MLC positioning errors and + 1 mm MLC gap width errors could be detected by the patient-specific VMAT QA procedure.The AD-DTA criteria of 2%/2 mm was more sensitive compared with the criteria of 3%/3 mm.Conclusions Patient-specific VMAT QA is not sensitive enough to detect the systematic MLC positioning errors within 1 mm.Additional MLC QA is needed to guarantee the accuracy of VMAT delivery.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2013年第4期388-391,共4页 Chinese Journal of Radiological Medicine and Protection
基金 国家自然科学基金(10975187)
关键词 容积调强旋转放疗 计划验证 多叶准直器 位置误差 通过率 Volumetric-modulated arc therapy ( VMAT ) Patient-specific QA Multileaf collimator (MLC) Positioning errors Passing rate
  • 相关文献

参考文献11

  • 1Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys, 2008, 35 (1) : 310-317.
  • 2Palma DA, Verbakel WF, OTTO K, et al. New developments in arc radiation therapy: a review. Cancer Treat Rev, 2010, 36 (5): 393-399.
  • 3Van Esch A, Huyskens DP, Behrens CF, et al. Implementing RapidArc into clinical routine: a comprehensive program from machine QA to TPS validation and patient QA. Med Phys, 2011, 38(9) : 5146-5166.
  • 40' DanielJ, Das S, W u QJ, et al. Volumetric-modulated arc therapy: effective and efficient end-to-end patient-specific 'quality assurance. IntJ Radiat Oncol Bioi Phys, 2012, 82 ( 5) : 1567 -1574.
  • 5马阳光,张可,胡志辉,宋一昕,戴建荣.利用加速器日志文件验证容积旋转调强放疗计划的可行性[J].中华放射医学与防护杂志,2012,32(3):285-288. 被引量:7
  • 6Yan G, Liu C, Simon TA, et al. On the sensitivity of patient?specific IMRT QA 10 MLC positioning errors.J Appl Clin Med Phys, 2009,10(1) : 2915.
  • 7Rangel A, Palte G, Dunscombe P. The sensitivity of patient specific IMRT QC to systematic MLC leaf bank offset errors. Med Phys, 2010, 37(7) : 3862-3867.
  • 8Garcia- Vicente F, Fernandez V, Bermudez R, et al. Sensitivity of a helical diode array device to delivery errors in IMRT treatment and establishment of tolerance level for pretreatment QA.J Appl Clin Med Phys, 2012,13(1) :3660.
  • 9Oliver M, Gagne I, Bush K, et al. Clinical significance of multi?leaf collimator positional errors for volumetric modulated arc therapy. Radiother Oncol, 2010, 97 (3) : 554-560.
  • 10Jursinic PA, Nelms BE. A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery. Med Phys, 2003, 30 (5) : 870-879.

二级参考文献13

  • 1Litzenberg DW,Moran JM,Fraass BA. Verification of dynamic and segmental IMRT delivery by dynamic log file analysis[J].Journal of Applied Clinical Medical Physics,2002,(02):63-72.
  • 2Stell AM,Li JG,Zeidan OA. An extensive log-file analysis of step-and-shoot intensity modulated radiation therapy segment delivery errors[J].Medical Physics,2004,(06):1593-1602.doi:10.1118/1.1751011.
  • 3Dinesh Kumar M,Thirumavalavan N,Venugopal Krishna D. QA of intensity-modulated beams using dynamic MLC log files[J].Medical Physics,2006,(01):36-41.
  • 4Schreibmann E,Dhabaan A,Elder E. Patient-specific quality assurance method for VMAT treatment delivery[J].Medical Physics,2009,(10):4530-4535.
  • 5Tony T,Bergman AM,Kwa W. Monte Carlo based,patient-specific RapidArc QA using linec log files[J].Medical Physics,2010,(01):116-123.doi:10.1118/1.3266821.
  • 6Sun Nuclear Corporation. ArcCheck reference guide[M].2010.
  • 7Low DA,Harms WB,Mutic S. A technique for the quantitative evaluation of dose distributions[J].Medical Physics,1998,(05):656-661.
  • 8ELEKTA. AutoCAL 2.0 User Manual[M].2008.1-36,91-118.
  • 9Gloi AM,Buchanan RE,Zuge CL. RapidArc quafity assurance through MapCHECK[J].J Appl Cliu Med Phys,2011,(02):39-47.
  • 10lftimia I,Cirino ET,Xiong L. Quality assurance methodology for Varian RapidArc treatment plans[J].Journal of Applied Clinical Medical Physics,2010,(04):130-143.

共引文献6

同被引文献172

引证文献19

二级引证文献107

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部