低心率下第二代双源CT冠脉成像Flash扫描与序列扫描图像质量及伪影比较 被引量:4

Comparing the image quality of coronary artery and the artifacts in lower heart rate between high-pitch and sequential scan mode on the second-generation dual-source CT
下载PDF
导出
摘要 目的:比较低心率下CT前瞻性心电触发Flash扫描与序列扫描的冠状动脉图像质量及伪影。方法:回顾性连续收集Flash扫描(A组)与序列扫描(B组)冠状动脉CT成像各80例患者,心率均≤70次/min。4级法评价2组图像质量,初步分析伪影特点及成因。结果:A、B 2组可评价冠状动脉节段为98.37%和99.29%,差异无统计学意义(χ2=3.56,P=0.094);可评价患者为85.00%和96.25%,差异有统计学意义(χ2=4.708,P=0.035);平均节段质量评分为(1.295±0.60)分和(1.21±0.59)分,差异有统计学意义(Z=-2.591,P=0.010)。A组的4级图像大部分分布于右冠中远段,多数出现在心率大于65次/min的患者中,均为心脏搏动伪影。B组4级图像散在分布于冠脉各段中,为呼吸及心脏搏动伪影。A组辐射剂量明显低于B组,差异有统计学意义。结论:低心率下Flash与序列扫描均可获得高质量冠状动脉图像,序列扫描略优于Flash扫描;Flash扫描右冠状动脉中远段出现心脏搏动伪影概率大,序列扫描易产生呼吸及心律不齐所致伪影。 Objective:To compare the image quality and analyze the artifacts between the prospective ECG-gated high-pitch spiral mode(Flash mode)and the prospective ECG-gated sequential acquisition mode(sequential mode)for coronary artery imaging in lower heart rate.Methods:Each group had 80subjects with heart rate≤70bpm.In group A,CT was performed using Flash mode,and the data acquisition was obtained at 60% of the R-R interval.In group B,Data was collected using the sequential mode at 35%-75% of the R-R interval.Image quality(IQ)was evaluated using a four-point scale(1=excellent,4= unevaluable).Results:There was no significant difference between two groups on the basis of segment in image quality(group A,98.37%;group B,99.29%;P=0.094),but there was significant difference on the basis of patient(group A,85%;group B,96.25%;P=0.035)and in mean IQ score between groups(1.295±0.60vs.1.21±0.59,Z=-2.591,P=0.010).In group A,all of IQ score in 4th were shown at the middle and distal segments of the right coronary artery,and these artifacts were not caused by respiratory motion,and the heart rates of most of these cases were more than 65bpm.In group B,however,the artifacts with IQ score in 4th were randomly distributed in three branches of coronary artery.They were caused by both of breathing and cardiac motion.Conclusion:At the low heart rate,the high quality of coronary artery imaging can be similarly acquired by Flash and sequential scan mode,respectively.However,in Flash mode,the cardiac motion artifacts happened more easily at the middle and distal segments of the right coronary artery.
出处 《中国中西医结合影像学杂志》 2013年第4期359-362,共4页 Chinese Imaging Journal of Integrated Traditional and Western Medicine
关键词 体层摄影术 X线计算机 冠状血管造影术 辐射剂量 质量控制 Tomography, X-ray computed Coronary angiography Radiation dose Quality control
  • 相关文献

参考文献13

  • 1Leschka S,Stolzmann P,Desbiolles L,et al.Diagnostic accuracyof high-pitch dual-source CT for the assessment of coronary ste-noses:first experience[J].Eur Radiol,2009,19:2896-2903.
  • 2王怡宁,李烁,孔令燕,王志伟,周慷,曹剑,樊红苓,张晓娜,张竹花,金征宇.低剂量双源CT前瞻性心电触发大螺距扫描冠状动脉造影的图像质量和放射剂量[J].中国医学科学院学报,2010,32(6):597-600. 被引量:21
  • 3赵青,佟海滨,王小玲,薛超,张民,杨树华,王群,汪平,石凤祥,徐磊,于薇,方继良.第二代双源CT冠状动脉成像前门控大螺距扫描的初步经验[J].中国中西医结合影像学杂志,2011,9(4):299-302. 被引量:3
  • 4Austen WG,Edwards JE,Frye RL,et al.A reporting system on pa-tients evaluated for coronary artery disease.Report of the Ad HocCommittee for Grading of Coronary Artery Disease,Council on Car-diovascular Surgery,American Heart Association[J].Circulation,1975,51:5-40.
  • 5Achenbach S,Anders K,Kalender WA.Dual-source cardiac com-puted tomography:image quality and dose considerations[J].EurRadiol,2008,18:1188-1198.
  • 6Achenbach S,Marwan M,Ropers D,et al.Coronary computedtomography angiography with a consistent dose below 1mSv u-sing prospectively electrocardiogram-triggered high-pitch spiralacquisition[J].Eur Herat J,2010,31:340-346.
  • 7佟海滨,王小玲,赵青,张民,薛超,杨树华,王群,汪平,石凤祥,徐磊,于薇,方继良.第二代双源CT前门控FLASH扫描与序列扫描冠状动脉成像比较[J].中国中西医结合影像学杂志,2011,9(5):400-404. 被引量:13
  • 8Blankstein R,Shah A,Pale R,et al.Radiation dose and imagequality of prospective triggering with dual-source cardiac compu-ted tomography[J].Am J Cardiol,2009,103:1168-1173.
  • 9Ketelsen D,Thomas C,Werner M,et al.Dual-source computedtomography:estimation of radiation exposure of ECG-gated andECG-triggered coronary angiography[J].Eur J Radiol,2010,73:274-279.
  • 10Achenbach S,Marwan M,Schepis T,et al.High-pitch spiral ac-quisition:a new scan mode for coronary CT angiography[J].JCardiovasc Comput Tomogr,2009,3:117-121.

二级参考文献49

  • 1Einstein AJ,Henzlova MJ,Rajagopalan S.Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography[J].JAMA,2007,298(3):317-323.
  • 2Lell M,Marwan M,Schepis T,et al.Prospectively ECGtriggered high-pitch spiral acquisition for coronary CT angiography using dual source CT:technique and initial experience[J].Eur Radiol,2009,19(11):2576-2583.
  • 3Achenbach S,Marwan M,Ropers D,et al.Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered highpitch spiral acquisition[J].Eur Heart J,2010,31(3):340-346.
  • 4Pflederer T,Jakstat J,Marwan M,et al.Radiation exposure and image quality in staged low-dose protocols for coronary dual-source CT angiography:a randomized comparison[J].Eur Radiol,2010,20(5):1197-1206.
  • 5Austen WG,Edwards JE,Frye RL,et al.A reporting system on patients evaluated for coronary artery disease.Report of the Ad Hoe Committee for Grading of Coronary Artery Discase,Council on Cardiovascular Surgery[J].Circulation,1975,51(4 Suppl):5-40.
  • 6Scheffel H,Alkadhi H,Plass A,et al.Accuracy of dualsource CT coronary angiography:first experience in a high pre-test probability population without heart rate control[J].Eur Radiol,2006,16(12):2739-2747.
  • 7Petersilka M,Bruder H,Krauss B,et al.Technical principles of dual source CT[J].Eur J Radiol,2008,68(3):362-368.
  • 8Achenbach S,Marwan M,Schepis T,et al.High pitch spiral acquisition:a new scan mode for coronary CT angiography[J].J Cardiovasc Comput Tomogr,2009,3(2):11-121.
  • 9Achenbach S,Marwan M,Ropers D,et al.Coronary computed tomography angiography with a consistent dose below 1mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition[J].Eur Heart J,2010,31:340-346.
  • 10Bamberg F,Marcus R,Sommer W,et al.Diagnostic image quality of a comprehensive high-pitch dual-spiral cardiothoracic CT protocol in patients with undifferentiated acute chest pain[J].Eur J Radiol,2010,31.

共引文献60

同被引文献18

引证文献4

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部