摘要
建立了水产品中毒杀芬残留量的凝胶渗透色谱净化-气相色谱检测方法。水产样品经匀浆后,用丙酮提取,经正己烷液液萃取、凝胶渗透色谱和浓硫酸净化后,用HP-5ms毛细管柱分离,电子捕获检测器检测,外标法定量。结果表明,毒杀芬在20~500μg/L浓度范围内线性良好,相关系数月≥0.998。方法检出限1.0μg/kg,在添加量2.0~20.0μgkg范围内,回收率在73.7%-78.7%之间,相对标准偏差≤6.3%。该方法灵敏、准确,可以满足水产品中毒杀芬检测要求。
A new method containing gel permeation chromatography (GPC) and gas chromatography with electron - capture detection ( GC - ECD) for the determination of toxaphene in aquatic products has been developed. After hom- ogenization, the samples were extracted by acetone, and then purified by liquid - liquid extraction. A further purifica- tion was conducted on a GPC system followed by a thorough clean - up with H2SO4. The separation was performed on an HP -5ms capillary column. Identification and quantification were conducted using GC -ECD. Toxaphene concen- tration exhibited a good linearity over the range of 20 -500 μg/L with R≥0.998. The detection limit of this method was 1.0 μg/kg for toxaphene. The recovery rates of toxaphene were 73.7% -78.7% at the range of 2.0 -20.0 μg/kg. This accurate and sensitive method can be applied for the determination of toxaphene residues in aquatic products.
出处
《中国渔业质量与标准》
2013年第3期73-77,共5页
Chinese Fishery Quality and Standards
基金
山东省现代农业产业技术体系刺参产业创新团队项目
国家海洋公益性行业科研专项(201105013)
山东省科学技术发展计划项目(2012GHY11517)
烟台市科技发展计划项目(2012134)
关键词
气相色谱
毒杀芬
水产品
残留
电子捕获检测器
gas chromatography ( GC )
toxaphene
aquatic product
residue
electron capture detector(ECD)