期刊文献+

维拉帕米选择性电极的研制和应用 被引量:4

Preparation and Application of Veraparnil Selective Electrodes
下载PDF
导出
摘要 报告以四苯硼酸-维拉帕米离子缔合物作活性材料,制成了普通 PVC 膜型、铂丝涂(PVC)膜型、碳棒涂(PVC)膜型和铂丝涂双层聚合膜型等各类维拉帕米电极,比较了它们的性能。在铂丝表面先修饰一层聚苯胺薄膜,继涂含活性材料的 PVC 膜所制成的双膜电极,线性范围为1.0×10^(-2)~5.7×10^(-6)mol/L,平均斜率为58.0mV/pC,检测下限为2.5×10^(-6)mol/L,在 pH3.0~6.5范围内适用。电极抗干扰能力强,响应快,重现性和稳定性较好。 Using verapamil-tetraphenylborate ion-pair complex as active material,four types of verapamil selective electrodes were prepared such as conventional PVC membrane, Pt-wire coated PVC membrane,carbon-rod coated PVC membrane and Pt-wire coated bilayer polymeric films.Platinum wire is modified by conductive polyaniline,and then coated by PVC membrane containing electroactive material.This kind of electrode shows Nernstian range from 1×10^(-2)mol/L to 5.7×10^(-6)mol/L with a slope of 58.0 mV/pC at pH 3.0~6.5, and the detection limit is 2.5×10^(-6)mol/L.The electrode has satisfactory selectivity,fast response,good reproducibility and stability.
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1991年第11期1301-1304,共4页 Chinese Journal of Analytical Chemistry
基金 江苏省教委自然科学基金
关键词 维拉帕米 离子选择电极 戊脉安 Verapamil Ion-selective electrode Coated wire electrode Coated carbon rod electrode Modified electrode
  • 相关文献

参考文献5

二级参考文献5

  • 1张丽珠,分析化学,1987年,15卷,12期,1081页
  • 2黄德培,化学传感器,1986年,6卷,1期,8页
  • 3沈国励,药物分析杂志,1984年,4卷,4期,200页
  • 4俞汝勤,分析化学,1983年,11卷,5期,343页
  • 5冷宗周,胡效亚.PVC阿米替林选择性电极的研究[J].分析化学,1989,17(2):139-142. 被引量:5