期刊文献+

两种房水引流植入装置降眼压效果比较 被引量:3

Effects of intraocular pressure drop for two kinds of aqueous humor drainage implant devices
原文传递
导出
摘要 目的观察并比较两种房水引流装置植入术降眼压的效果。方法选择健康新西兰白兔40只,单眼进行房水引流装置植入术。根据不同的植入物,分为NADI组和Ahmed组,每组20只实验兔。结果术后各组均未出现浅前房。NADI组术后8周眼压比术前下降19.4%;Ahmed组术后8周眼压下降幅度为15.7%。术后8周内,各组间眼压值统计对比,均有显著性差异(P<0.05),且NADI组的眼压值下降幅度要明显大于Ahmed组。结论 NADI组比Ahmed组的眼压值下降幅度大,且下降时间维持较长,表明此新型房水引流植入物NADI降眼压效果明显,具有很好的临床应用前景。 Objective To observe and compare the effects of intraocular pressure (IOP) drop on using two kinds of aqueous drainage device implantation. Methods Forty healthy New Zealand white rabbits were chosen for the experiments. According to different implant devices, they were divided into NADI (new aqueous drainage implant) group and Ahmed group respectively, with 20 rabbits in each. Results Neither group appeared shallow anterior chamber after operation. At 8th week after operation, the postoperative IOP of NADI group and Ahmed group was decreased by 19.4% and 15.7%, respectively, as compared to the preoperative IOP. Statistical comparison showed that there were significant differences in IOP between two groups within 8 weeks (P〈0.05), and the IOP of NADI group was decreased more significantly than that of Ahmed group. Conclusions The IOP in NADI group was reduced more than that in Ahmed group, and such IOP drop was maintained for a longer period in NADI group, which means that NADI has an obvious IOP drop effect and good prospect in clinical application.
出处 《医用生物力学》 EI CAS CSCD 北大核心 2014年第1期66-71,共6页 Journal of Medical Biomechanics
基金 国家自然科学基金资助项目(31070840) 北京市属市管高等学校人才强教深化计划项目高层次人才项目(PHR201110506) 首都医科大学基础临床合作课题重点项目(12JL03)
关键词 青光眼引流装置 房水 引流 眼压 Glaucoma drainage devices(GDD) Aqueous humor Drainage Intraocular pressure (IOP)
  • 相关文献

参考文献8

  • 1郭学谦,田蓓,孙世杰,谢楠,刘志成.高眼压对青光眼视网膜功能的影响[J].医用生物力学,2010,25(3):195-199. 被引量:5
  • 2Minckler DS, Francis BA, Hodapp EA, et al. Aqueous shunts in glaucoma: A report by the American Academy of Ophthalmology [J]. Ophthalmology, 2008, 115 (6) : 1089- 1098.
  • 3Johnson B, House P, Morgan W, et al. Developing laser- induced glaucoma in rabbits [J]. Aust N Z J Ophthalmol, 1999, 27(3-4) : 180-181.
  • 4The Krupin eye valve filtering surgery study group. Krupin eye valve with disk for filt surgery [ J ]. Ophthalmology, 1994, 101(4) : 651-654.
  • 5Nguyen AH. Avoiding and managing complications of glau- coma drainage implants E J ]. Curr Opin Ophthalmol, 2004, 15(2) : 147-150.
  • 6马红.青光眼小梁切除术后浅前房原因及治疗[J].医药论坛杂志,2006,27(10):46-47. 被引量:6
  • 7Huang MC, Netland PA, Coleman AL, et al. Intermediate- term clinical experience with the Ahmed Glaucoma Valve implant [J]. Am J Ophthalmol, 1999, 127(1): 27-33.
  • 8Yalvac IS, Eksioglu U, Satana B, et al. Long-term results of Ahmed glaucoma valve and Molteno implant in neovascu- lar glaucoma [J]. Eye, 2007, 21(1): 65-70.

二级参考文献17

  • 1视觉电生理国际标准化委员会.视网膜电图国际标准,1989.
  • 2Gil BS, Sharon B, George NL, et aL Pattem electroretinography in a rat model of ocular hypertension: functional evidence for early detection of inner retinal damage [ J ]. Experimental Eye Research, 2005, 81(9):340-349.
  • 3Thom WM, John D, Geralyn P, et aL Retinal damage after 3 to 4 months of elevated intraocular pressure in a rat glaucoma model[J]. Invest Ophthalmol Vis Sci, 2000, 41 (10) : 3451-3459.
  • 4Sinisa DG, Young HK, Donald SS, et al. Functional evaluation of retina and optic nerve in the rat model of chronic ocular hypertension [ J ]. Exp Eye Res, 2004, 79 (7) : 75- 83.
  • 5Lijun J, William OC, Elaine C J, et al. Effect of general anesthetics on lOP in rats with experimental aqueous outflow obstruction[Jl. Invest Ophthalmol Vis Sci,2000, 41 (11) : 3415-3419.
  • 6Balwantray CC, Jinyi P, Michele LA, et aL Effect of intraocular pressure on optic disc topography, electroretinography, and axonal loss in a chronic pressure-induced rat model of optic nerve damage [ J ]. Invest Ophthalmol Vis Sci, 2002, 43(9) ; 2969-29?6.
  • 7Bayer AU, Danias J, Brodie S, et al. Electroretinographic abnormalities in a rat glaucoma model with chronic elevated intraocular pressure[J]. Exp Eye Res, 2001, 72(6):667- 677.
  • 8Brad F, Bang VB, Tohn CM, et al. Selective ganglion cell functional loss in rats with experimental glaucoma[J]. Invest Ophthalmol Vis Sci, 2004, 45 (6) : 1854-1862.
  • 9Yoko I, Jacky MKK, Joseph C, et al. Retinal ganglion cell protection with geranylgeranylacetone, a heat shock protein inducer, in a rat glaucoma model [ J ]. Invest Ophthalmol Vis Sci. , 2003, 44(5 ) : 1982-1992.
  • 10Solon Thanos, Rita Naskar. Correlation between retinal ganglion cell death and chronically developing inherited glaucoma in a new rat mutant[ J]. Exp Eye Res, :2004, 79 (7) :119-129.

共引文献9

同被引文献26

引证文献3

二级引证文献2