期刊文献+

免疫放射分析与微粒子化学发光法测定血清TSH的价值 被引量:1

Immunoradiometric and chemiluminescent assays for determination of serum thyrosin stimulating hormone
原文传递
导出
摘要 目的:观察免疫放射分析与微粒子化学发光法检验血清促甲状腺素(TSH)的价值。方法:分别采用免疫放射分析与微粒子化学发光法检测40份血清标本的TSH水平。结果:免疫放射法、微粒子化学分析法测量的TSH水平分别为(7.889±0.355)、(8.077±0.299)μIU/mL,两者比较,差异无统计学意义(t=1.532,P>0.05)。两种方法测定TSH值具有较好的相关性(γ=0.99)。结论:免疫放射分析与微粒子化学发光法检验血清TSH水平具有较好相关性。 Objective: To determine the diagnostic value of immunoradiometric and chemiluminescence assay for assessment of serum thyrosin stimulating hormone (TSH). Methods: Forty serum samples were subject to measurement of TSH by using the two different techniques. Results: Immunoradiometric and chemiluminescence assay yielded an unmarked difference in the detected level of TSH [ ( 7. 889 ±0. 355 ) vs (8. 077 ± 0. 299) μIu/mL, P 〉 0. 05 ]. The measured TSH values correlated well between the two techniques ( r = 0.99) Conclusion: The measured TSH values correlated well between the two techniqugs.
作者 冯玉俊
出处 《广州医学院学报》 2013年第5期67-68,共2页 Academic Journal of Guangzhou Medical College
关键词 免疫放射分析法 微粒子化学发光法 血清TSH immunoradiometric assay method chemiluminescence technique serum thyrosin stimulatinghorlnone
  • 相关文献

参考文献4

  • 1杨昌国,许叶,张抗.精密度评价和方法比较中NCCLS评价方案的应用[J].临床检验杂志,1999,17(1):47-49. 被引量:184
  • 2Spencer CA,Takeuchi M.Current status and performance goals for serum thyrotropin(TSH) assay[J].Clin chem,1996,42(1):140-145.
  • 3Gosling JPA.Decade of development in immunoassodology[J].Clin Chem,1990,56(8):1408-1427.
  • 4OgnibeneA,Drark CJ,Jeng KY,et al.A new modular chemiluminescence immunoassay analyser evaluated[J].Clin Chem lab Med,2000,38(3):251-260.