大体积进样离子色谱法对电厂水汽中痕量阴离子测定中两种分离柱的分离效果 被引量:7

Efficiency of 2 Different Separation Columns in IC Determination of Traces of Anions in Steam of Power Generating Unit with Large Volume Sample Introduction
下载PDF
导出
摘要 对两种阴离子分离柱(AS 17-C和AS 19)在大体积进样(500 μL)离子色谱法测定发电机组蒸汽中8种痕量阴离子中的分离效果做了试验和比较.现场取得的水样经滤膜过滤后直接进样,分别经上述两柱分离,并用不同浓度的氢氧化钾溶液作为流动相进行梯度淋洗,用抑制电导检测器检测.8种阴离子的质量浓度均在5~50 μg·L-1范围内呈线性,检出限(3S/N)分别在0.07~0.73 μg·L-1,0.08~1.9μg·L-1之间.用标准加入法做回收试验,测得回收率在99.6%~104%(AS 17-C柱)和98.0%~103%(AS 19柱)之间. The efficiency of 2 different anion separation columns (i.e.,AS 17-C and AS 19) used in the IC determination of 8 trace anions with large volume sample introduction (500 μL) was studied comparatively.The water sample taken from worksite was filtered on membrane and introduced directly into the instrument.Separation of the 8 anions was performed separately on the 2 columns.KOH solutions with various concentrations were used as mobile phase in the gradient elution,and inhibitory conductivity detection was adopted in IC analysis.Linearity for the 8 anions was found same in the range of 5 50 μg · L-1,while the detection limits (3S/N) found were in the ranges of 0.07-0.73 μg · L-1 and 0.08-1.90 μg · L-1 for the 2 columns respectively.Values of recovery found by standard addition method were in the ranges of 99.6%-104% (for AS 17 C) and 98.0%-103% (for AS 19).
出处 《理化检验(化学分册)》 CAS CSCD 北大核心 2014年第1期101-104,共4页 Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
关键词 离子色谱法 阴离子分离柱 8种痕量阴离子 大体积进样 发电机组蒸汽 IC Anion separation column 8 Trace anions Large volume sample introduction Steam of power generating unit
  • 相关文献

参考文献5

二级参考文献12

共引文献3

同被引文献68

引证文献7

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部