期刊文献+

ICP-AES法同时测定氯化锂和氢氧化锂中七种杂质元素 被引量:17

Simultaneous Determination of Seven Impurity Elements in Lithium Chloride and Lithium Hydroxide by ICP AES
下载PDF
导出
摘要 本文报导了用ICP-AES法同时测定LiCl和LiOH·H2O中7种杂质元素Al、Ba、Ca、Fe、Mg、Si、Zn的分析方法。研究了基体元素锂对被测元素的基体效应,采用基体匹配法与背景扣除法进行校正。被测元素的检出限为0.1-9.4ng/mL,加标试验回收率为92%-109%,当杂质元素含量为0.0001%-0.028%时,相对标准偏差小于7%。方法简便、快速、准确,用于样品分析,取得了满意的结果。 In this paper, a method has been studied for simultaneous determination of seven impurity elements Al , Ba , Ca , Fe , Mg , Si and Zn in LiCl and LiOH·H 2O by ICP AES. Matrix effect from lithium to the spectral lines of the elements determined were investigated and corrected by the matrix maching and background correction method. The detection limits of every element are 0.1-9.4ng/mL, The recoveries are 92%-109%, the relative standard deviations are less than 7% for impurity elements content 0.0001%-0.028%. This method is simple, rapid and accurate, and has been applied to the analysis of sample with satisfactory results.
出处 《光谱实验室》 CAS CSCD 1998年第3期39-42,共4页 Chinese Journal of Spectroscopy Laboratory
  • 相关文献

参考文献1

  • 1金泽祥,郑毅.ICP发射光谱法的干扰效应[J]分析试验室,1988(01).

同被引文献737

引证文献17

二级引证文献78

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部