期刊文献+

TOC/TN测定仪燃烧氧化电化学法测定水中的总氮

Determination of total nitrogen in water by burningoxidation electrochemical methodusing TOC/TN analyzer
下载PDF
导出
摘要 建立了采用TOC/TN测定仪燃烧氧化电化学法测定水中总氮的分析方法,优化了仪器最佳测定参数。方法检出限≤0.05mg·L^(-1),线性范围宽0~200mg·L^(-1),实际样品相对标准偏差<4%,加标回收率96.0%~103.0%。该方法无需另加试剂和前处理、快速简便、无干扰,适用于环境监测中各种水质中总氮的测定。 An analytical method for the determination of total nitrogen in water by burning oxidation electrochemical method using TOC/TN analyzerwas established,and the optimal measurement parameters of the instrument were optimized.The detection limit of the method was ≤0.05mg·L^-1,the linear range was 0-200mg·L^-1,the relative standard deviation of the actual sample were <4%,and the recovery rate were 96.0%-103.0%.The method does not require additional reagents and pretreatment,quick and easy,no interference,and was suitable for the determination of total nitrogen in various water quality in environmental monitoring.
作者 徐义邦 张吴 陈芬 XU Yi-bang;ZHANG Wu;CHEN Fen(Nanchang Environmental Monitoring Station,Nanchang 330038,China)
出处 《化学工程师》 CAS 2018年第12期26-28,共3页 Chemical Engineer
关键词 总氮 TOC/TN测定仪 燃烧氧化电化学法 紫外光度法 total nitrogen TOC/TN analyzer,burning oxidation electrochemical method ultraviolet spectrophotometry
  • 相关文献

参考文献6

二级参考文献18

  • 1史竞男,刘荣梅.测定总氮时应注意的几个问题[J].环境科学与管理,2005,30(4):97-98. 被引量:6
  • 2赵虹.水质总氮测定中影响因素分析[J].辽宁城乡环境科技,2005,25(6):12-14. 被引量:3
  • 3中华人民共和国国家标准.GB11894-89.水质-总氮的测定-碱性过硫酸钾消解紫外分光光度法[S].北京:中国标准出版社,1989:192-195.
  • 4余艳华 吴志旭 徐李琦.环境样品监测中总氮小于氨氮问题探讨.环境污染与防治,2008,(5):1-2.
  • 5水和废水监测分析方法(总氮紫外分光光度法)[M].第4版.北京:中国环境科学出版社,2002.
  • 6国家环境保护总局.水和废水监测分析方法[M].第四版.北京:中国环境科学出版社,2002年12月.
  • 7刘建宁.总氮测定值偏低的原因浅谈及解决[J].黄山市环境监测站,2012:1008-1151.
  • 8张宪伟.2012.国标法测定总氮过程中若干常见问题的探讨[J].科学观察,(3): 64-65.
  • 9HJ636-2012,水质总氮的测定碱性过硫酸钾消解紫外分光光度法[S].北京:环境科学出版社,2012.
  • 10国家环境保护总局,水和废水监测分析方法编委会编,魏复盛.水和废水监测分析方法[M]中国环境科学出版社,2002.

共引文献56

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部