两种HIV-1 RNA定量检测方法的临床对比研究 被引量:7

Comparison of two assays for HIV-1 RNA quantitation
原文传递
导出
摘要 目的比较达安基因艾滋病病毒1型(HIV-1)核酸检测Ⅱ代(简称达安基因)和罗氏Cobas TaqMan HIV-1Test Version 2.0(简称罗氏Cobas TaqMan)两种试剂,检测HIV-1血浆病毒载量的相关性及一致性。方法采用达安基因试剂和罗氏Cobas TaqMan试剂,平行检测663份HIV-1感染者血浆病毒载量,采用配对t检验、相关分析和Bland-Altman等方法进行统计分析。结果罗氏Cobas TaqMan与达安基因试剂检测结果高于检测下限的比例一致(分别为65.31%和65.76%)。对于病毒载量低于1 000IU/mL的样本,二者的差异无统计学意义(P=0.856 8)。罗氏Cobas TaqMan与达安基因试剂检测结果具有很强的相关性(r=0.962 0,P<0.000 1)。此外,Bland-Altman分析两种方法检测病毒载量的结果具有较高的一致性。结论达安基因和罗氏Cobas TaqMan试剂,检测HIV-1血浆病毒载量的结果具有较好的相关性和一致性。 Objective To compare DaanGene’s human immunodeficiency virus type 1(HIV-1)test version 2.0(DaanGene)reagent and Roche’s Cobas TaqMan HIV-1test version 2.0(Cobas TaqMan)for HIV-1plasma viral load quantitation,and investigate the correlation and consistency of the two assays.Methods The plasma viral load of 663HIV-1infected patients was detected in parallel by DaanGene and Cobas TaqMan assays.The results obtained from the two assays were analyzed using paired t test,correlation analysis and Bland-Altman.Results The proportion of HIV-1plasma viral load above the lower limit of detection determined by Cobas TaqMan and DaanGene was consistent(65.31%and 65.76%,respectively).For samples with viral load less than 1 000IU/mL,there was no significant difference between the two assays(P=0.856 8).In addition,HIV-1plasma viral load detected by Cobas TaqMan was strongly correlated with that by DaanGene’reagent(r=0.962 0,P<0.000 1).Furthermore,Bland-Altman analysis showed that there was good consistency for the viral loads obtained from the three assays.Conclusion There are great correlation and consistency between Roche’s Cobas TaqMan and DaanGene’s HIV-1test reagents for HIV-1plasma viral load quantitation.
作者 蔡侃儒 程林 陈伟梅 黄乙甘 唐娴 徐六妹 何云 王辉 CAI Kanru;CHENG Lin;CHEN Weimei;HUANG Yigan;TANG Xian;XU Liumei;HE Yun;WANG Hui(Shenzhen Third People's Hospital,Shenzhen 518112, China)
出处 《中国艾滋病性病》 CAS CSCD 北大核心 2019年第1期18-21,共4页 Chinese Journal of Aids & STD
基金 国家自然科学基金(31500750) 广东省自然科学基金(2016A030313030) 深圳市科技创新基金(JCYJ20160427183909738 JCYJ20170412151650600) 深圳市三名工程(SZSM201512029)~~
关键词 艾滋病病毒1型 病毒载量 相关性 一致性 HIV-1 Viral load Correlation Consistency
  • 相关文献

参考文献5

二级参考文献40

  • 1潘品良,蒋岩.HIV-1病毒载量测定及质量保证指南(试行)[S].中国疾病预防控制中心.2008:13-16.
  • 2Mellors JW, Munoz A, Giorgi JV, et al. Plasma viral load and CD4 lymphocytes as prognostic markers of HIV-1 infeetion[J]. Ann Intern Med, 1997, 126(12): 946--954.
  • 3Bland JM, Altman DG. Statistical methods for assessing agree- ment between two methods of clinical measurement[J]. Lancet, 1986,1 (8476) : 307 --310.
  • 4中国疾病预防控制中心.全国艾滋病检测技术规范[S].200126-27.
  • 5Gomes P, Carvalho AP, Diogo I, et al. Comparison of the Nu-eliSENS EasyQ HIV-1 v2.0 with Abbott m2000rt RealTimeHIV- 1 assay for plasma RNA quantitation in different HIV-1 subtypes [J]. Journal of Virological Methods, 2013, 193(1): 18--22.
  • 6Bourlet T, Signori-Schmuck A, Roche L, et al. HIV-1 Load Comparison Using Four Commercial Real-Time Assaysl[J]. Jour- nal of Clinical Microbiology, 2011, 49 (1) : 292-- 297.
  • 7Lyles CM, Vlahov D, Farzadegan H, et al. Comparison of Two Measures of Human Immunodefieieney Virus(HIV) Type 1 Load in HIV Risk Groups[J]. Journal of clinical microbiology, 1998, 36(12) :3647--3652.
  • 8Damond F, Roquebert B, Benard A, et al. Human im/nunodefi- ciency virus type 1 (HIV-1) plasma load discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR Version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 assays [J]. J Clin Microbiol, 2007, 45(10): 3436--3438.
  • 9Wolff D, Gerritzen A. Comparison of the Roche COBAS Ampli- eor Monitor,Roche COBAS Ampliprep/COBAS Taqman and Ab- bott RealTime Test assays for quantification of hepatitis C virus and HIV RNA[J]. Clin Chem Lab Med, 2007, 45(7).. 917--922.
  • 10Robertson K, Fiscus S, Kapoor C, etal. CSF, plasma viral load and HIV associated dementia[J]. Journal of NeuroVirology, 1998, 4(1) 90--94.

共引文献190

同被引文献43

引证文献7

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部