摘要 目的:探讨XT-1800i血细胞分析仪Q-Flag报警信息与血涂片复检的相符情况以及在临床中的应用。方法:收集笔者所在医院2013年6-12月所有Q-Flag报警信息标本,并对其进行分析。结果:Q-Flag≤120 U时仪器与人工血涂片复检符合率为0,Q-Flag 120~220 U时仪器与人工涂片复检符合率为1.5%~72.9%,Q-Flag 220~300 U时,仪器与人工涂片复检符合率为88.9%~100%。结论:Q-Flag≤120 U人工涂片复检无实际意义,Q-Flag 120~300 U,人工涂片复检在疾病的诊断、治疗和疗效观察中起重要作用。 Objective:To investigate the situation between the Q-Flag alarm information of XT-1800i hematology analyzer and blood smear reexamination and their clinical significance. Method:Collected all samples of Q-Flag alarm from June 2013 to December 2013 in the hospital. Result:When it alarmed Q-Flag≤120 U,the coincidence rate was zero by manual smear. Q-Flag 120-220 U,the coincidence rate was 1.5%-72.9%. Q-Flag 220-300 U,the coincidence rate was 88.9%-100%. Conclusion:It’s no practical significance when Q-Flag≤120 U,when Q-Flag 120-300 U,reexamination of blood smear plays an important role in disease diagnosis,treatment and the observation of curative effect.
出处 《中外医学研究》 2014年第20期76-78,共3页 CHINESE AND FOREIGN MEDICAL RESEARCH
关键词 Q-Flag报警信息 血涂片复检 符合率 Q-Flag alarm information Reexamination of blood smear Coincidence rate
  • 相关文献

参考文献10

二级参考文献43

共引文献47

同被引文献24

  • 1Xiong W,Ong S H,Lim J H,et al.Automatic Area Classification in Peripheral Blood Smears[J].IEEE Transactions on Biomedical Engineering,2010,57(8):1982-1990.
  • 2Ghosh M,Chakraborty C,Konar A,et al.Development of hedge operator based fuzzy divergence measure and its application in segmentation of chronic myelogenous leukocytes from microscopic image of peripheral blood smear[J].Micron,2014,57(2):41-55.
  • 3Prasad K,Winter J,Bhat U M,et al.Image analysis approach for development of a decision support system for detection of malaria parasites in thin blood smear images[J].Journal of Digital Imaging,2012,25(4):542-549.
  • 4Etienne C L,Granat F,Trumel C,et al.A mycobacterial coinfection in a dog suspected on blood smear[J].Veterinary Clinical Pathology,2013,42(4):516-521.
  • 5Ikegaya S,Tai K,Shigemi H,et al.Fulminant candidemia diagnosed by prompt detection of pseudohyphae in a peripheral blood smear[J].Am J Med Sci,2012,343(5):419-420.
  • 6Lee J,Lee S Y,Yoo S S,et al.Clinical value of whole-blood interferon-gamma assay in patients with suspected pulmonary tuberculosis and AFB smear and polymerase chain reactionnegative bronchial aspirates[J].Diagnostic Microbiology and Infectious Disease,2012,73(3):252-256.
  • 7唐仕华,何星.血涂片铁染色在贫血诊断中的应用[J].中国医学创新,2010,7(1):172-172. 被引量:2
  • 8乐家新,丛玉隆,王海,王红霞,傅淑宏,徐菡,兰亚婷,朱红.Sysmex XE-2100血细胞分析仪血涂片复检规则的应用研究[J].现代检验医学杂志,2010,25(1):30-36. 被引量:31
  • 9伏攀,李春碧.用新鲜全血校准血细胞分析仪的方法[J].中国医学创新,2010,7(6):130-132. 被引量:4
  • 10王馥香.三分类血细胞分析仪检测中间型细胞镜检的探讨[J].中国医学创新,2010,7(9):160-161. 被引量:1

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部