摘要
目的:通过滤波技术,排除震荡电位干扰,准确测量闪光视网膜电图a波、b波的隐含期及振幅。方法:对15例30眼门诊健康体检者行闪光视网膜电图检查,通频带为0.6~300Hz和0.6~70Hz时,分别测量a波、b波的隐含期及振幅,并进行对比分析,采用配对t检验。结果:将通频带设置为0.6~70Hz时,a波、b波表现为单峰,对比通频带为0.6~300Hz时,a波、b波的隐含期延长,振幅下降(P〈0.01)。结论:将通频带设置为0.6~70Hz时能获得光滑的a波、b波波形,能准确测量a波、b波的隐含期及振幅。
AIM: To accurately measure the implicit time and amplitude of a-, b-wave from flash electroretinogram ( ERG) through filtering technology, eliminate oscillatory potentials interference. METHODS: Full-field ERGs were recorded in 30 eyes of 15 physical check-ups, measured the implicit time and amplitude of a-, b-wave, when the passband was set at 0. 6 - 300Hz and 0. 6 - 70Hz, and correlation was performed among those results by paired t-test.-RESULTS: When the passband was set at 0.6 - 70Hz, a-, b-wave had a single peak, compared with the passband was set at 0.6 - 300Hz, the implicit time of a-, b-wave was prolonged, amplitude was decreased ( P〈0.01 ). CONCLUSION : A passband of 0.6 - 70Hz was the best choice to obtain smooth a- and b-waves from the original ERG, It could accurately measure the implicit time and amplitude of a-, b-wave.
出处
《国际眼科杂志》
CAS
2014年第11期2031-2033,共3页
International Eye Science
关键词
视网膜电图
视网膜
震荡电位
electroretinogram
retina
oscillatory potentials