期刊文献+

微波消解-ICP/MS动态反应池技术测定汽油中的磷 被引量:3

Microwave digestion ICP-MS dynamic reaction cell technology determination of phosphorus in gasoline
下载PDF
导出
摘要 建立电感耦合等离子体质谱法(ICP-MS)测定汽油中磷的分析方法,样品经过微波消解,利用ICP-MS动态反应池(DRC)技术,选择高纯氧气作为反应气,使P+转化为PO+,确定了仪器的最佳参数和实验的最佳条件。实验表明:氧气流量在0.65 m L/min、Rpq值在0.55时,磷元素在10.0μg/L^100.0μg/L呈良好线性,相关系数为0.999 9,RSD<1.5%,检出限为0.4μg/L,加标回收率在93%~107%,能够满足日常汽油中磷的测定。 Analysis method of ICP-MS determination of phosphorus in gasoline, samples after microwave digestion, using ICP-MS dynamic reaction cell(DRC)technology, selection of high pure oxygen as the reactive gas, convert P+to PO+, the optimal conditions for optimal parameters and experimental instrument determination. Experimental results show, the flow of oxygen in the 0.65 m L/min and Rpq values in 0.55, phosphorus is linear in 10.0 μg/L ~100.0 μg/L, the correlation coefficient of 0.999 9, RSD〈1.5 %, the detection limit is 0.4 μg/L,recovery of standard addition is 93 %~107 %, can meet the daily determination of phosphorus in gasoline.
出处 《石油化工应用》 CAS 2014年第11期118-120,123,共4页 Petrochemical Industry Application
关键词 微波消解 电感耦合等离子体质谱法 动态反应池技术 汽油 microwave digestion inductively coupled plasma mass spectrometry dynamic reaction cell technology gasoline phosphorus
  • 相关文献

参考文献9

二级参考文献20

共引文献28

同被引文献24

引证文献3

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...