摘要 选用IonPac AS19阴离子分析柱,以KOH为淋洗液梯度淋洗,对大体积进样直接测定饮用水中痕量溴酸盐的离子色谱方法进行了条件优化。结果表明,在优化好的色谱条件(即电流量87mA、淋洗液流速为1.00mL/min、梯度淋洗)下,在BrO3^-为2.0~100.0μg/L的低浓度范围内,峰面积与浓度呈良好的线性关系(r=0.999 6),测定的精密度高。采用该方法测定市售瓶装水样品中的痕量BrO3^-,样品的平均加标回收率为95.91%~105.50%。该方法操作简单,BrO3^-的分离效果好,可与常见阴离子实现同时分析,且测定精密度高,可作为饮用水中痕量溴酸盐的测定方法。 To optimize a large volume direct injection ion chromatography method for conductivity detection of bromate in drinking water,the IonPac AS19 column with KOH as gradient eluent was selected and used.The results showed that under the optimal chromatographic conditions(electric current 87 mA,leacheate's flow rate 1.00mL/min,gradient elution),the correlation coefficient between peak area and bromate concentration was 0.999 6when the bromate ranged from 2.0to 100.0μg/L.The method was used for the analysis of trace bromate in commercially available bottled water,with the average recoveries were between 95.91%-105.50%.The method was simple,accurate,sensitive and applicable for the determination of trace bromate in drinking water.
出处 《环境污染与防治》 CAS CSCD 北大核心 2015年第1期72-75,共4页 Environmental Pollution & Control
关键词 离子色谱 溴酸盐 饮用水 ion chromatography bromate drinking water
  • 相关文献

参考文献16

二级参考文献51

  • 1史亚利,蔡亚岐,刘京生,牟世芬,温美娟.大体积直接进样离子色谱法测定饮用水中痕量溴酸盐[J].分析化学,2005,33(8):1077-1080. 被引量:52
  • 2郭强.饮用水液氯消毒机理及其消毒副产物的控制[J].山西科技,2005(5):111-112. 被引量:7
  • 3臧道德,童俊.柱后反应-离子色谱法测定饮用水中的溴酸盐[J].净水技术,2005,24(5):66-68. 被引量:7
  • 4臧道德,童俊,冯菊丽,陈国光,刘肖,牟世芬.淋洗液在线发生离子色谱法测定水中的溴酸盐[J].中国给水排水,2006,22(18):83-85. 被引量:7
  • 5侯小平,王玉敏,等.第七届全国色谱学术报告会论文集,1999:62-64.
  • 6[1]Determination of Inorganic Anions in Water by Ion Chromatography.U.S.Environmental Protection Agency Method 300.1,Cincinnati,Ohio,1997
  • 7[2]Determination of Inorganic Oxyhalide Disinfection By-roducts in Drinking Water Using Ion Chromatography with the Addition of a Postcolumn Reagent for Trace Bromate Analysis.U.S.Environmental Protection Agency Method 317.2,Cincinnati,Ohio,2001
  • 8[3]Determination of Bromate in Drinking Waters by Ion Chromatography Inductively Coupled Plasma-mass Spectrometry.U.S.Environmental Protection Agency Method 321.8,Cincinnati,Ohio,1997
  • 9[4]Wagner H P,Pepich B V,Hautman D P et al.,Eliminating the Chlorite Interference in US Environmental Protection Agency Method 317.0 Permits Analysis of Trace Bromate Levels in All Drinking Water Matrices.J.Chromatogr.A,2000,882:209-319
  • 10Richardson S D, Thruston A M, Caughran T V. DBPs in Drinking Water. Proc. -Water Qual. Technol. Conf. 1998.