铅眼镜对介入放射工作人员眼晶状体的防护效果研究 被引量:12

Shielding effect of lead glasses on eye lens exposure in interventional radiologists
原文传递
导出
摘要 目的 探讨不同铅当量和镜片大小铅眼镜对介入放射工作人员眼晶状体的防护效果.方法 利用MCNPX软件和人体体素体模,模拟计算在不同投照体位条件下介入放射学工作人员佩戴不同铅眼镜时的眼晶状体剂量;通过在仿真人体模上布放眼晶状体剂量计开展实验测量,以验证理论计算结果.结果 模拟计算结果表明,当单个镜片大小为20 cm2时,佩戴0.1~1.0mmPb(铅)眼镜可降低眼晶状体受照剂量系数为3~9;剂量降低系数(DRF)不仅与眼镜的铅当量有关,而且还取决于X射线的投照体位;当镜片的铅当量超过0.35 mm Pb时,随着镜片铅当量的增加,剂量的进一步降低并不明显.在相同铅当量条件下,当镜片尺寸为6~30 cm2时,DRF与镜片的大小成线性增长关系.模拟计算数值和实验结果基本吻合.结论 为有效地保护介入放射学工作人员的眼晶状体,推荐佩戴铅当量为0.5 mm Pb且镜片尽量大的防护眼镜. Objective To study the shielding effect of equivalent thickness and size of lead glasses on eye lens of interventional radiologists.Methods Based on the human voxel phantom and MCNPX software,doses to eye lens of radiologists were simulated under different conditions of wearing lead glasses and beam projections,and measurements were performed with anthropomorphic phantom placed with eye lens dosimeters to verify simulation results.Results The results showed that the dose to eye lens reduced by a factor from 3 to 9 when the equivalent thickness of glasses ranged from 0.1 to 1.0 mm Pb with a len size of 20 cm2.The dose reduction factors (DRF) not only depended on the lead equivalent,but also on the beam projection of X-rays.However,the increase in DRF was not significant whenever the lead equivalent of glasses was larger than 0.35 mm.Furthermore,the DRF was proportional to the size of glass lens from 6 to 30 cm2 with the same lead equivalent.The simulation results were in well agreement with the measured ones.Conclusions For more reasonable and effective protection of the eye lens of interventional radiologists,a pair of glasses with a lead equivalent of 0.5 mm and large-sized lens is recommended.
机构地区 [
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2014年第12期927-930,共4页 Chinese Journal of Radiological Medicine and Protection
基金 国家自然科学基金(11075040)
关键词 铅眼镜 介入放射学 眼晶状体 剂量 蒙特卡罗模拟方法 Lead glasses Interventional radiology Eye lens Dose Monte Carlo simulation methods
  • 相关文献

参考文献18

  • 1Balter S.Stray radiation in the cardiac catheterization laboratory[J].Radiat Prot Dosim,2001,94(1-2):183-188.
  • 2International Commission on Radiological Protection.ICRP Publication 89.Avoidance of radiation injuries from medical interventional procedures[R].Oxford:Pergamon Press,2000.
  • 3Farah J,Struelens L,Dabin J,et al.A correlation study of eye lens dose and personal dose equivalent for interventional cardiologists[J].Radiat Prot Dosim,2013,157(4):561-569.
  • 4McVey S,Sandison A,Sutton DG.An assessment of lead eyewear in interventional radiology[J].J Radiol Prot,2013,33(3):647-659.
  • 5Antic V,Ciraj-Bjelac O,Rehani M,et al.Eye lens dosimetry in interventional cardiology:results of staff dose measurements and link to patient dose levels[J].Radiat Prot Dosim,2013,154(3):276-284.
  • 6李农,王良群,缪维芳.唐山市放射工作人员眼晶体损害调查分析[J].中国辐射卫生,2011,20(3):383-383. 被引量:3
  • 7薛娴,岳保荣,尉可道.介入放射学职业人员眼晶状体剂量的测量与评估[J].中华放射医学与防护杂志,2013,33(1):108-110. 被引量:8
  • 8Carinou E,Ferrari P,Koukorava C,et al.Monte Carlo calculations on extremity and eye lens dosimetry for medical staff at interventional radiology procedures[J].Radiat Prot Dosim,2011,144(1-4):492-496.
  • 9International Commission on Radiological Protection.ICRP Publication 110.Adult reference computational phantoms[R].Oxford:Pergamon Press,2007.
  • 10Bilski P,Bordy JM,Daures.J,et al.The new EYE-DTM dosemeter for measurements of HP (3) for medical staff[J].Radiat Meas,2011,46(11):1239-1242.

二级参考文献4

共引文献9

同被引文献136

  • 1白雪岭,刘谦,余雳,廖银萍,骆清铭,龚辉.适于小动物研究的断层解剖成像系统[J].航天医学与医学工程,2006,19(1):62-65. 被引量:5
  • 2曾志,李君利,邱睿,贾向红.质子剂量微分谱预估空间辐射剂量[J].清华大学学报(自然科学版),2006,46(3):374-376. 被引量:8
  • 3卫生部令第55号,放射工作人员职业健康管理办法[S].
  • 4GBZ128-2002.职业性外照射个人监测规范[S].[S].,..
  • 5GB18871-2002.电离辐射防护与辐射源安全基本标准[S].[S].,..
  • 6放射工作人员职业健康管理办法[J].首都公共卫生,2007,1(5):191-194. 被引量:10
  • 7Xu XG, Chao TC, Bozkurt A. VIP-Man: An image-based whole- body adult male model constructed from color photographs of the visible human project for multi-particle Monte Carlo calculation[J]. Health Phys, 2000,78: 476-486.
  • 8Saito K, Wittmann A, Koga S, et al. Construction of a computed tomographic phantom for a Japanese male adult and dose calc- ulation system[J]. Radiat Environ Biophys, 2001, 40: 69-76.
  • 9Nagaoka T, Watanabe S, Sakurai K, et al. Development of realistic high resolution whole-body voxel models of Japanese adult males and females of average height and weight and application of models to radio-frequency electromagnetic field dosimetry[J]. Phys Med Biol, 2004, 49: 1-15.
  • 10Lee C, Lee J, Lee C. Korean adult male voxel model Korman segmented from magnetic resonance images[J]. Med Phys, 2004, 31 (5): 1017-1022.

引证文献12

二级引证文献46

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部