摘要
目的:探讨应用低温等离子体灭菌器对十二指肠镜灭菌效果及成本、有效期的分析研究。方法:收集急诊科使用过的十二指肠镜,分别采用低温等离子体灭菌器与常规灭菌器进行消毒,比较两种方法的灭菌效果、成本及有效期。结果:低温等离子体灭菌器每次灭菌成本为(78.21±7.57)元,常规灭菌器每次灭菌成本为(44.82±5.26)元;低温等离子体灭菌器灭菌有效期为5~7个月,常规灭菌器灭菌有效期为9~28 d。结论:低温等离子体灭菌法具有灭菌力强、低温、快速及有效期长等优势,值得在临床推广及应用。
Objective: To Study on the application of low temperature plasma sterilizer for endoscopic disinfection and cost, period of validity. Methods: Used duodenum mirrors were collected in December 2012-December 2013 in our department, respectively by low temperature plasma sterilizer with conventional disinfection sterilizer. To compare two methods of sterilization effect and cost, the period of validity. Results: Low temperature plasma sterilizer every time sterilization costs RMB(78.21+7.57), significantly higher than the conventional sterilizer(44.82 5.26) yuan(t=7.71, P0.05); Low temperature plasma sterilizer sterilization is 5 to 7 months, longer than conventional sterilizer 9-28 d(t=12.50, P〈0.05). Conclusion: Low-temperature plasma sterilization method has strong sterilization, temperature, rapid and long shelf life and other advantages, worthy of promotion and application.
出处
《中国医学装备》
2015年第2期51-53,共3页
China Medical Equipment