DXI800化学发光仪测定cTnI出现假阳性结果的分析和处理

Analysis and treatment on false positive result of cTnI detection by DXI800 Chemiluminescence Apparatus
下载PDF
导出
摘要 目的:分析和处理15例DXI800全自动化学发光分析仪检测的肌钙蛋白I(c Tn I)假阳性结果,以降低假阳性率,提高检验结果的准确性。方法:用DXI800全自动化学发光分析仪进行c Tn I测定,对结果≥0.1 ng/m L的标本进行复查。结果:共检测10 529人次的c Tn I,结果异常(〉0.04 ng/m L)1 137人次,其中766人次结果≥0.1 ng/m L。对≥0.1 ng/m L的结果进行复查,复查后出现15例为0.01 ng/m L(原结果为0.31-2.52 ng/m L)。在排除仪器、试剂、室内质控原因后确定为假阳性。结论:标本前处理环节非常重要,分离的血清标本是否合格都有可能引起结果出现偏差,造成假阳性或假阴性结果,应严格按检验前处理要求进行标本处理。 Objective To analyze and treat the 15 false positive results of c Tn I detection by DXI800 Chemiluminescence Apparatus. Methods c Tn I detection was performed with DXI800 Chemiluminescence Apparatus, and the specimens with the results not less than 0.1 ng/m L were rechecked. Results Among 10 529 c Tn I results, there were 1 137 abnormal ones(0.04 ng/m L) which included 766 ones not less than 0.1 ng/m L. Recheck showed that there were 15 results of 0.01 ng/m L,with the original results between 0.31 and 2.52 ng/m L. Then 15 false positive results were found after the factors of instrument, reagent and indoor quality control excluded. Conclusion Improper pretreatment may cause false positive or negative results, and thus the specimen has to be pretreated according to the requirements.
出处 《医疗卫生装备》 CAS 2015年第4期86-88,共3页 Chinese Medical Equipment Journal
关键词 肌钙蛋白I 假阳性 DXI800化学发光仪 cTnI false positive DXI800 Chemiluminescence Apparatus
  • 相关文献

参考文献7

二级参考文献73

  • 1郭新胜,高元法,张秀明,高耀东,周卫强,郭艳蕊,张迎黎,姜锋,吕惠娟.分离胶处理标本不适合离子选择电极法监测血锂浓度[J].临床检验杂志,2004,22(4):288-289. 被引量:3
  • 2汪俊汉,张艳平.劣质血清分离胶对甘油三酯测定的影响[J].国际检验医学杂志,2006,27(9):847-848. 被引量:5
  • 3Universal definition of myocardial infarction. European Heart Journal, 2007,28 : 2525-2538.
  • 4[2]蒋秉坤主编.临床生物化学及生物化学检验.北京:人民卫生出版社,1996.5-14.
  • 5[3]寇丽筠主编.临床基础检验学.北京:人民卫生出版社,1996.280-290.
  • 6Pedoe-Tunstall H,Kuulasmaa K, Amouyel P,et al. Myocardial infarction and coronary deaths in the World Health ()rganization MONICA Project.Circulation, 1994,90: 583.
  • 7Goldberg R,Gore J,Alpert J,et al. Incidence and case fatality tates of acute myocardial infarction( 1975 - 1984 ) : The Worcester Heart Attack Study. Am Heart J, 1988,115 : 761.
  • 8Kannel W. Prevalence and clinical aspects of unrecognized myocardial infarction and sudden unexpected death. Circulation, 1987,75(Suppl Ⅱ) :Ⅱ.
  • 9Gibler W, Lewis L, Erb R, et al. Early detection of acute myocardial infarction in patients presenting with chest pain and nondiagnostic ECGs:Serial CKMB sampling in the emergency deparment. Am Emerg Med,1990,19: 1359.
  • 10Adams JE, Bodor GS, Davlia-Roman VG, et al. Cardiac troponin I: A marker with high specificity for cardiac injury. Circulation, 1993,88:101.

共引文献60

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部