125I粒子植入前装载过程中医护人员的剂量监测 被引量:5

Monitoring of radiation dose to medical staff during 125I seeds loading
原文传递
导出
摘要 目的 分析1255I粒子装载过程中医护人员的受照剂量.方法 采用热释光剂量计检测125I装载过程中,与粒子槽不同距离的剂量,计算出人员的安全范围;医护人员的受照剂量、剂量率和年累积剂量.结果 与粒子槽不同距离处10、20、30、40、50和100 cm的剂量衰减率分别是77.61%、98.04%、98.79%、99.30%、99.71%和100%.医护人员的指尖、胸部、眼晶状体和甲状腺的受照剂量分别是51.08、35.50、34.73和33.78 μGy,年操作250次达到的剂量分别是12.77、8.88、8.68和8.45 mGy.铅玻璃屏内、外粒子剂量的衰减率分别是28.36%、79.60%.结论 125I粒子装载过程中,安全距离超过100 cm时检测不到辐射,铅玻璃防护是很有必要的. Objective To analyze the dose of radiation to medical staff during 125I seeds loading Methods The radiation dose at different distances was measured by using thermoluminescence dosimeters (TLD),and the safe range for medical person was calculated.The doses and annual accumualive doses to the medical staff were estimated.Results The dose attenuation rates were 77.61%,98.04%,98.79%,99.30%,99.71% and 100% at distance of 10,20,30,40,50 and 100 cm from seeds tank,respectively.The dose to fingertips,chest,lens and thyroid of doctor were 51.08,35.50,34.73 and 33.78 μGy,and the annual dose reached 12.77,8.88,8.68 and 8.45 mGy when the number of annual operations was 250.The attenuation rate was 79.60% and 28.36% of inside and outside lead glass,respectively.Conclusions The radiation is undetectable at 100 cm away from seeds tank in the process of loading of 125I seeds,and lead glass is necessary for radiation protection.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2015年第7期536-538,共3页 Chinese Journal of Radiological Medicine and Protection
基金 首都医学发展科研基金(2009-2024)
关键词 125I粒子 热释光剂量计 剂量监测检测 125I seeds Thermoluminescence dosimeter Dose surveillance
  • 相关文献

参考文献12

  • 1Meng J, Wang X, Zhuang Qw, et al. Clinical effectiveness of t25i_seed implantation in combination with nimotuzumab therapy for the advanced oral carcinoma: preliminary results [ J ]. Eur Rev Med Pharmacol Sci, 2014, 18 (21) : 3304-3310.
  • 2Beydoun N, Bucci J, Malouf D. Iodine-125 prostate seed brachytherapy in renal transplant recipients: an analysis of oncologica| outcomes and toxicity profile [ J ]. J Contemp Brachytherapy, 2014, 6( 1 ) : 15-20.
  • 3Sekiguchi A, Ishiyama 14, Satoh T, et a|. |odine-125 monotherapy for Japanese men with low- and intermediate-risk prostate cancer : outcomes after 5 years of follow-up [ J ]. J Radiat Res, 2014, 55(2) : 328-333.
  • 4Bhengu K J, Langen KM, Symons JE, et al. Thermoluminescence dosimetry in a fast neutron therapy beam [J]. Radiat Phys Chem, 2004, 71: 981-982.
  • 5国家质量监督检验检疫总局.GB18871-2002电离辐射防护与辐射源安全基本标准[S].北京:中国标准出版社,2002.
  • 6van Haaren PM, van't Riet A, Moerland MA, et al. Dose to fingertips of staff preparing stranded iodine-125 seeds for permanent prostate implants[J]. Radiat Prot Dosim, 2011, 145 (1): 61-65.
  • 7耿建华,曹仲年,赵卫强,洪浩,邹作伟,孙超,郑容,陈盛祖.^(125)Ⅰ粒子装载过程中工作人员受照剂量的研究[J].中国医学装备,2012,9(10):1-3. 被引量:10
  • 8Magee JS, Martin CJ, Sandblom V, et al. Derivation and application of dose reduction factors for protective eyewear worn in interventional radiology and cardiology [ J ]. J Radiol Prot, 2014, 34(4): 811-823.
  • 9岳瑶.北京市2287名放射体检人员信息调查分析[J].中国职业医学,2009,36(2):177-177. 被引量:9
  • 10岳瑶,王文学,梁莉.某医院114名放射工作人员健康调查分析[J].中国辐射卫生,2008,17(3):308-309. 被引量:8

二级参考文献35

共引文献82

同被引文献44

引证文献5

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部