根管预备前后Raypex 6根管长度测量仪的测量准确性研究

Clinical Application of Raypex 6 Endometer for Root Canal Treatment
下载PDF
导出
摘要 目的探讨Raypex 6根管长度测量仪在根管预备前后测量根管工作长度的准确性。方法选择因正畸或者修复拔除的单根管离体牙齿100颗,常规开髓,根管预备前使用插针X线片法以及长度测量仪法分别测量根管工作长度,根管预备至K#30号后使用长度测量仪测量工作长度,比较根管预备前X线片法和测量仪法的差异性以及根管预备后根管长度测量仪的准确性。结果根管预备前,X线片法与Raypex 6根管长度测量仪法所测得的根管工作长度分别为(18.505±1.006)mm和(18.440±1.026)mm,两组比较无显著性差异(t=2.944,P=0.04);根管扩大后Raypex 6根管长度测量仪所测得的根管工作长度为(17.615±1.939)mm,与根管预备前X线片法所测得的(18.505±1.006)mm比较有显著性差异(t=5.313,P=0.000)。结论 Raypex 6根管长度测量仪是一种准确的测量工作长度的方法,但根管扩大会降低Raypex 6根管长度测量仪的准确性。 Objective To evaluate the accuracy of Raypex 6 before and after root canal preparation.Methods A total of 100 single rooted teeth with mature apices after extracted were selected for this study. The distance of actual working length was recorded using X-ray methods and using apex locators. After the preparation root canals reached to K#30,the distance of electronic apex locators was recorded. Results Before root canal preparation,the root canal length measured by the X-ray method and Raypex 6 was( 18. 505 ± 1. 006) mm vs.( 18. 440 ±1. 026) mm. There was no significant difference between the two groups( t =2. 944,P = 0. 04). While comparing with the X-ray method,the post- preparation root canal length measured by Raypex 6 was significantly shorter,which was( 17. 615 ± 1. 939) mm vs.( 18. 505 ± 1. 006) mm( t = 5. 313,P = 0. 000). Conclusion Raypex 6 is an accuracy method but the accuracy would decrease after root canal preparation.
作者 蒋红 孔令佳
出处 《中国现代手术学杂志》 2015年第3期234-236,共3页 Chinese Journal of Modern Operative Surgery
关键词 Raypex 6 根管长度测量仪 根管介质 根尖孔 Raypex 6 electronic root canal length measurement device root canal media apical foramen
  • 相关文献

参考文献5

  • 1Ahenburger M J, Tchorz JP, Somma F. In vivo accuracy of three e- lectronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II[ J-. Int Endod J, 2012,45 ( 11 ) : 1053- 1054. doi: 10. 1111/j. 1365-2591. 2012. 02060. x.
  • 2Aydin U, Karataslioglu E, Aksoy F, et al. In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters[ J]. J Conserv Dent, 2015,18( 1 ) :66-69. doi: 10.4103/0972-0707. 148899.
  • 3Akisue E, Gratieri SD, Barletta FB, et al. Not all electronic fora- men locators are accurate in teeth with enlarged apical foramina: an in vitro comparison of 5brands [ J ]. J Endod, 2014 ,40 ( 1 ) : 109-112. doi: 10. 1016/i. ioen. 2013.09. 032.
  • 4Moscoso S, Pineda K, Basilio J, et al. Evaluation of Dentaport ZX and Raypex 6 electronic apex locators : an in vivo study [ J ]. Med Oral Patol Oral Cir Bucal, 2014,19(2) :e202- e205.
  • 5徐义华,孟健.探讨RootZX根管长度测量仪使用中准确性及影响因素[J].中国现代医生,2009,47(25):179-180. 被引量:3

二级参考文献5

共引文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部