期刊文献+

一种基于胸阻抗法的心功能无创检测分析仪 被引量:1

A kind of heart function analytical device based on impedance cardiography method
下载PDF
导出
摘要 目的为了经济、快捷、全面地实现心功能的无创检测,基于胸阻抗法成功研发了一种心功能无创检测分析仪,可方便地实现胸阻抗信号、心电信号、心音信号的同步检测分析,从而实现对患者心功能的无创综合评价。该方法无毒无创,操作简单,完全可以实现家用化普及。方法本文首先描述了系统的硬件模块构成,说明了胸阻抗信号的采集过程。其次,使用FPGA芯片与DDS芯片构成系统的控制与信号发生核心,指出了恒流源的精度等性能指标。再次,指出了胸阻抗信号处理的要点,运用互感原理实现干扰信号的隔离。最后,介绍了仪器软件功能,并展示了仪器软件实测结果。结果通过临床试验,对胸阻抗法与超声多普勒法检测的数据结果进行t检验,结果表明二者具有一致性。结论由于采用了先进的特征点判别方法,该设备具有较高的临床检测精度和较好的临床适用性,可满足临床心功能无创检测和评估的要求。 Objective A kind of heart function analytical device is invented based on impedance cardiography method to detect heart function economically,fast and comprehensively. The device can detect and analyze ICG,ECG and PCG signals synchronously,so comprehensive noninvasive assessment of heart function in patients can be achieved. Methods This article introduces the system and main hardware modules of the device, describes the process of ICG acquisition. FPGA and DDS are used as hardware core and performance indices of the device are pointed out. Key points of signal processing are presented and the electromagnetic principle of mutual inductance is used in this part. Results By analyzing and comparing clinical data,this article indicates that the device can gain more accurate clinical test data and has better clinical applicability. Conclusions Because of its advanced feature points' extraction method,this device meets the requirements of clinical examination and assessment for heart function.
出处 《北京生物医学工程》 2015年第5期489-494,共6页 Beijing Biomedical Engineering
基金 国家自然科学基金(81371713)资助
关键词 胸阻抗 仪器设计 临床数据分析 chest impedance medical device design clinical data analysis
  • 相关文献

参考文献15

  • 1Bernstein DP, Lemmens HJM. Stroke volume equation for impedance cardiography[J]. Medical and Biological Engineering and Computing, 2005, 43(4): 443-450.
  • 2Gujjar AR, Muralidhar, K, Banakal S, et al. Non-invasive cardiac output by transthoraeie electrical bioimpedence in post- cardiac surgery patients: comparison with thermodilution method [J]. Journal of clinical monitoring and computing, 2008, 22 (3) : 175-180.
  • 3Fellahi JL, Caille V, Charron C, et al. Noninvasive assessment of cardiac index in healthy volunteers: a comparison between thoracic impedance cardiography and Doppler echocardiography [J]. Anesthesia & Analgesia, 2009, 108(5) : 1553-1559.
  • 4Kamath SA, Drazner MH, Tasissa G, et al. Correlation of impedance cardiography with invasive hemodynamic measurements in patients with advanced heart failure: the bioimpedance cardiography (BIG) substudy of the evaluation study of congestive heart failure and pulmonary artery catheterization effectiveness ( ESCAPE ) trial [ J ]. American Heart Journal, 2009, 158(2) : 217-223.
  • 5Hauck S, DeHon A. Reconfigurable computing: the theory and practice of FPGA-based computation [ M ]. San Francisco: Morgan Kaufmann, 2010.
  • 6Bour J, Kellett J. Impedance cardiography a rapid and cost- effective screening tool for cardiac disease [ J ]. European Journal of Internal Medicine, 2008, 19(6) : 399-405.
  • 7Tostevin F, Ten Wolde PR. Mutual information between input and output trajectories of biochemical networks [ J ]. Physical Review Letters, 2009, 102(21): 218101.
  • 8Bernstein DP. Impedance cardiography: pulsatile blood flow and the biophysical and eleetrodynamie basis for the stroke volume equations[ J]. Journal of Electrical Bioimpedanee, 2010, 1 ( 1 ) : 2-17.
  • 9宋义林,高树枚,五十岚朗,山越宪一.基于电阻抗法无拘束心排量检测的最佳电极配置[J].生物医学工程学杂志,2011,28(1):32-35. 被引量:6
  • 10Patterson R P. Impedance cardiography: What is the source of the signal? Journal of Physics : Conference Series[ C ]. USA : IOP Publishing, 2010:112-118.

二级参考文献23

  • 1NYBOER J. Electrical impedance plethysmogrsphy: A physi- cal and physiological approach to peripheral vascular study [J]. Circulation, 1950, 2:811-821.
  • 2KUBICEK W G, PATERSON R P, WITSOE D A. Impedance cardiography as a noninvasive method of monitoring cardiac function and other parameters of the cardiovascular system [J]. Ann NY Acad Sci, 1970, 170; 724-732.
  • 3PENNEY B C, PATWARDHAN N A, WHEELER H B. Simplified electrode array for impedance cardiography [J]. Med Biol Eng & Comput, 1985, 23,1-7.
  • 4GOTSHALL R W, SEXSON W R. Comparison of band and spot electrodes for the measurement of stroke volume by the bioeleetrie impedance teehnlque [J], Critical Care Medicine, 1994, 22:420-425.
  • 5中川原実,山越憲一.電流分布計測に基づいた電気的イソビ一ダソス心拍出量計測用スボシト電極配置[J].医用電子と生体工学,2000,3:238-245.
  • 6IKARASHI A, SASAKI K, MOROE T, et al, Quantitative e- valuation of optimal electrode array for impedance electrieal cardiography [C]. 18th SICE Symposiumon Biological and Physiological Engineering, Nigata, 2003:377-378.
  • 7YAMAKOSHI Y, TANAKA S, SHIMAZU H. Electrical ad- mittance cuff for non-invasive and simultaneous measurement of hematocrit, arterial pressure and elasticity using volume-os- cillometric method [J]. Med & Biol Eng & Comput, 1994, 32, 99-107.
  • 8Tsadok S.The historical evolution of bioimpedanbce[J].AACN Clinical Issues,1999,10 (3):371-384.
  • 9Osypka MJ,Bernstein DP.Electrophysiologic princil\ples and theory of stroke volume determination by thoracic electrical bioimpedance[J].AACN Clinical Issues,1999,10 (3):385-399.
  • 10Sramek BB.Physiological chart for rapid identification of causes of abnormal haemodynamics[J].Ann Acad Med.1994,23 (6 suppl):26-32.

共引文献12

同被引文献1

引证文献1

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部