期刊文献+

重力输液器用作泵用输液器的评估 被引量:3

Evaluation of gravity infusion set integrated in infusion pump
下载PDF
导出
摘要 目的 :评估将重力输液器用作泵用输液器的可行性。方法 :参照现行的医用注射泵和输液泵校准规范,利用输液设备分析仪开展用量较大的4款输液泵调校前后的检测,并进行统计学分析。结果:直接将重力输液器用作输液泵耗材,输液流速偏差较大,而根据所用的重力输液器对输液泵进行调校后,偏差变小。结论:重力输液器用作泵用输液器会增加输液风险,须通过必要的质量检测手段,筛选比较后方可应用。 Objective To evaluate the feasibility to integrate the gravity infusion set into the infusion pump. Methods Four commonly-used infusion pumps were detected by the infusion apparatus analyzer with references to the current specifications for medical infusion set and infusion pump calibration, and then went through statistical analysis, Results Flow velocity deviation might be high when the gravity infusion set used as the consumable of the infusion pump, and went low after calibrating the infusion set.Conclosion Risks may occur when integrating the gravity infusion set into the infusion pump directly, and some quality detection measures have to be involved in to decrease the risks.
出处 《医疗卫生装备》 CAS 2016年第1期85-89,共5页 Chinese Medical Equipment Journal
关键词 输液泵 输液器 评估 infusion pump infusion set evaluation
  • 相关文献

参考文献7

  • 1国家食品药品监督管理总局.2012年医疗器械不良事件监测年度报告发布[EB/OL].(2013-04-03)[2015-05-05].http://www.sda.gov.cn/WS01/CL0051/79538.html.
  • 2国家食品药品监督管理总局.国家医疗器械不良事件监测年度报告(2013年度)[EB/OL].(2014-05-07)[2015-03-01].http://www.sda.gov.cn/WS01/CL0438/99420.html.
  • 3国家食品药品监督管理总局.医疗器械不良事件信息通报(2011年第4期)关注输液泵、注射泵输注速度异常的风险[EB/OL].(2011-07-29)[2015-03-01].http://www.sda.gov.cn/WS01/CL0438/64377.html.
  • 4邵蕾,杨坤,乐莹,郑博.专用输液泵管路临床应用价值的评估[J].中国医疗器械杂志,2012,36(4):302-303. 被引量:14
  • 5于树滨.医用输液泵、注射泵质量控制检测技术[M].北京:中国计量出版社,2010:13.
  • 6刘丽琼,张平,朱琼.输液泵临床应用风险及管理对策[J].医疗卫生装备,2014,35(6):96-98. 被引量:18
  • 7GB 9706127 2005/IEC 60601-2-24 1998医用电气设备第2-24部分输液泵和输液控制器安全专用要求[S].